User`s guide

Agilent EasyEXPERT User’s Guide Vol. 1, Edition 1 1- 101
Main GUI
Data Display
Setup Summary
This function is effective for the I/V Sweep and Multi Channel I/V Sweep classic
tests. If the Setup Summary ON is set on the Graph Properties dialog box or the
Data Display Properties dialog box, the channel setup summary information shown
below is displayed at the bottom of the graph area.
The setup for one channel is displayed with one line as follows. Where, the items in
{} are the setting values.
VAR1 Channel For the voltage output channel:
{V Name}: [{Start}, {Stop}] V / {No of Step}; {I Name}: {Compliance} A; {Pwr Comp} W
For the current output channel:
{I Name}: [{Start}, {Stop}] A / {No of Step}; {V Name}: {Compliance} V; {Pwr Comp} W
Example: VDS: [0, 10] V / 101; ID: 1 A; 1 W
VAR2 Channel For the voltage output channel:
{V Name}: [{Start}, {Stop}] V / {Step} V; {I Name}: {Compliance} A; {Pwr Comp} W
For the current output channel:
{I Name}: [{Start}, {Stop}] A / {Step} A; {V Name}: {Compliance} V; {Pwr Comp} W
Example: VGS: [0, 5] V / 1 V; ID: 1 mA; 1 mW
Constant Channel For the voltage output channel:
{V Name}: {Source} V; {I Name}: {Compliance} A
For the current output channel:
{I Name}: {Source} A; {V Name}: {Compliance} V
For the common channel:
{V Name}: COMMON
Example: VSUB: -1 V; ISUB: 1 mA