User`s guide
Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1 7-11
Utilities
Desktop EasyEXPERT
Calibration Differences on Calibration window
• Calibration window provides Module Self Calibration screen only.
• Module Self Calibration screen provides Enable Auto Calibration only.
Configuration,
Main Frame
Differences on Configuration window Main Frame screen
• Line Frequency just displays the setting of 4155/4156.
• Firmware Rev. displays as follows.
[HOSTC Rev.]:[SMUC Rev.]:[ADC Rev.]
• Main Frame Diagnosis is not supplied.
Configuration,
Module
Differences on Configuration window Module screen
• Self-test cannot be performed.
• Status, SCUU, and Notes columns are not supplied.
• Start Self Test and Recover Module buttons are not supplied.
• Accessory Configuration is not supplied.
I/V Sweep Differences on I/V Sweep Classic Test
• Time Stamp Name field is not supplied for 4155B/4156B.
• Number of sweep steps must be 1 to 1001 for VAR1.
• SMU Pulse Period and Width value must satisfy Period Width + 4 ms.
I/V-t Sampling Differences on I/V-t Sampling Classic Test
• Log sampling is not supported.
• Sampling interval must be 60 s to 65.535 s.
• Number of samples must be 1 to 10001.
• Base hold time must be 0 sec.
ADC and
Integration Time
Differences on A/D Converter & Integration Time Setup
• ADC always shows HR ADC.
• High Resolution ADC is set as shown in Table 7-2.
Integration Time is ignored for the sampling measurement of Interval < 2 ms.