7-10 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Utilities
Desktop EasyEXPERT
Using 4155B/4156B/4155C/4156C
This section describes the notices when using Desktop EasyEXPERT and
4155/4156.
4155/4156
Firmware Revision
4155/4156 firmware revision supported by Desktop EasyEXPERT
HOSTC: 03.08 or later
SMUC: 04.08 or later
4155/4156
Functions
4155/4156 functions unsupported by Desktop EasyEXPERT
Log sampling measurement and Thinned out sampling measurement
Functions using R Box
Differential voltage measurement using VMU (Voltage Monitor Unit)
DC bias output using PGU (Pulse Generator Unit)
Other functions unavailable for the B1500
Desktop
EasyEXPERT
Classic tests available for 4155/4156
I/V Sweep, I/V-t Sampling, and Switching Matrix Control for 4155C/4156C
I/V Sweep and Switching Matrix Control for 4155B/4156B
Desktop EasyEXPERT functions unavailable for 4155/4156
Tracer Test
Functions using SMU Series Resistor
Functions using ASU (Atto Sense Unit)
Functions using SCUU (SMU CMU Unify Unit)
Functions using CMU (Capacitance Measurement Unit)
Functions using SPGU (Semiconductor Pulse Generator Unit)
PGU can be used instead of SPGU for the I/V Sweep and I/V-t Sampling classic
tests. The Load Z function and the ALWG output are not supported. The Output
Impedance dialog box is available to set the output impedance of the PGU.
Functions using WGFMU (Waveform Generator/Fast Measurement Unit) and
RSU (Remote-sense and Switch Unit)