User`s guide
Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1 6- 19
Remote Control Interface
CALibration Subsystem
:CALibration[:SMU]:ZERO[:ON]:ALL
Sets the all SMU to the offset current cancel function ON.
Syntax :CALibration[:SMU]:ZERO[:ON]:ALL
Example :CAL:ZERO:ALL
:CALibration[:SMU]:ZERO:PLC
Sets or returns the integration time used for the offset current measurement of the
specified SMU.
Syntax :CALibration[:SMU]:ZERO:PLC "module", plc
:CALibration[:SMU]:ZERO:PLC? "module"
Parameter module Identifier of the SMU. Data type is SPD.
plc Integration time, 1|2|5|10|16|25|50|100. This is the number of
power line cycles (PLC). If the invalid value is specified, the
maximum allowable value less than the specified value is set
automatically. Data type is NR1
Query response plc<newline>
Data type is NR1.
Example :CAL:ZERO:PLC "SMU4:HR/AS", 25
:CAL:ZERO:PLC? "SMU4:HR/AS"
:CALibration[:SMU]:ZERO:STATe?
Returns the present status of the offset current cancel function ON/OFF.
Syntax :CALibration[:SMU]:ZERO:STATe?
Query response mode<newline>
0 (OFF) or 1 (ON) is returned. Data type is Bool.
Example :CAL:ZERO:STAT?