6- 18 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Remote Control Interface
CALibration Subsystem
Query response current1[,current2[,current3[,current4]]]<newline>
Data type is NRf.
current1: Offset current measurement data for the 1 nA range
current2: Offset current measurement data for the 100 pA range
current3: Offset current measurement data for the 10 pA range
current4: Offset current measurement data for the 1 pA range
Example :CAL:ZERO:MEAS
:CAL:ZERO:MEAS? "SMU4:HR/AS"
:CALibration[:SMU]:ZERO:OFF:ALL
Sets the all SMU to the offset current cancel function OFF.
Syntax :CALibration[:SMU]:ZERO:OFF:ALL
Example :CAL:ZERO:OFF:ALL
:CALibration[:SMU]:ZERO[:ON]
Sets the specified SMU to the offset current cancel function ON. Or returns the
identifier of the SMU set to the offset current cancel function ON.
Syntax :CALibration[:SMU]:ZERO[:ON] "module1"[,"module2"[, ... "module"]]
:CALibration[:SMU]:ZERO[:ON]?
Parameter module Identifier of the SMU to be set to the offset current cancel
function ON. Data type is SPD.
Query response "module1"[,"module2"[, ... "module"]]<newline>
Identifier of the SMU set to the offset current cancel function ON. Data type is SRD.
Example :CAL:ZERO "SMU1:HP", "SMU2:MP", "SMU3:MP"
:CAL:ZERO?