User`s guide

Agilent EasyEXPERT Users Guide Vol. 2, Edition 1 6- 17
Remote Control Interface
CALibration Subsystem
CALibration Subsystem
Describes CALibration subsystem commands used for the offset current cancel
function.
:CALibration[:SMU]:ZERO:FULLrange
Sets or returns the full range measurement ON/OFF status for the specified
SMU+ASU.
Syntax :CALibration[:SMU]:ZERO:FULLrange "module", mode
:CALibration[:SMU]:ZERO:FULLrange? "module"
Parameter module Identifier of the SMU. Data type is SPD.
mode Full range measurement ON or OFF, 0|OFF|1|ON. Data type is
Bool.
0 | OFF: Full range measurement OFF
1 | ON: Full range measurement ON
Query response mode<newline>
0 (OFF) or 1 (ON) is returned. Data type is Bool.
Example :CAL:ZERO:FULL "SMU4:HR/AS", 1
:CAL:ZERO:FULL? "SMU4:HR/AS"
:CALibration[:SMU]:ZERO:MEASure[:CURRent]
Starts the offset current measurement for the SMU set to the offset current cancel
function ON. Or returns the offset current measurement data of the specified SMU.
Use the
*OPC? command to confirm the completion of the measurement.
Syntax :CALibration[:SMU]:ZERO:MEASure[:CURRent]
:CALibration[:SMU]:ZERO:MEASure[:CURRent]? "module"
Parameter module Identifier of the SMU. Data type is SPD.