10-52 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Error Message
Agilent FLEX Execution Error
3309 Specified loop number of ALWG Sequence Data is out of absolute
limits.
Check the loop value of the sequence data and set the correct value.
3310 Specified output voltage of ALWG Waveform Data is out of absolute
limits.
Check the output voltage and set the correct value. The value must be
-3 V to +3 V for the 3 V range, -5 V to +5 V for the 5 V range, -10
V
to 0 V for the -10 V range, or 0 V to +10 V for the + 10 V range.
3311 Specified interval time of ALWG Waveform is out of absolute limits.
Check the incremental time (interval time) and set the correct value.
The value must be 10 ns to 10995.11627775 s, in 10 ns resolution.
3312 Specified ALWG measurement interval time is out of absolute limits.
Check the measurement interval time and set the correct value. The
value must be 10 ns to 1.34217728 s, in 10 ns resolution.
3313 Specified ALWG measurement instruction code is invalid.
Check the measurement event setting and set the correct values.
3314 Specified ALWG range change instruction code is invalid.
Check the range event setting and set the correct values.
3315 Specified ALWG measurement count is out of absolute limits.
Check the measurement averaging time and set the correct value. The
value must be 0, or 10 ns to 0.020971512 s, in 10 ns resolution.
3316 Specified ALWG measurement count is greater than measurement
interval.
Check the measurement averaging time and set the correct value. The
value must less than or equal to the measurement interval time.
3317 Specified slot is invalid.
Check the slot number and set the correct value. The slot number must
be 1 to 10.
3318 Specified module channel is invalid.
Check the channel number and set the correct value.