User`s guide
Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1 10-45
Error Message
Agilent FLEX Execution Error
694 DIO control mode must be HVSMU Current Expander control mode
(ERMOD 8).
Set the digital IO control mode to N1266A control mode by using the
ERMOD 8 command.
695 Mismatch HVSMU Current Expander ID.
HVSMU current expander is not connected to the Digital IO connector.
Connect the N1266A.
696 HVSMU Current Expander control cable was disconnected.
Connection cable was removed. Turn the instrument off and connect
the cable. And then turn the instrument on again.
697 HVSMU Current Expander is not active.
HVSMU current expander does not respond. The power switch may be
off or the power code may be removed.
940 DIO control mode must be Ultra High Voltage Expander control mode
(ERMOD 16).
Set the digital IO control mode to N1268A control mode by using the
ERMOD 16 command.
941 Mismatch Ultra High Voltage Expander ID.
Ultra high voltage expander is not connected to the Digital IO
connector. Connect the N1268A.
942 Ultra High Voltage Expander control cable was disconnected.
Connection cable was removed. Turn the instrument off and connect
the cable. And then turn the instrument on again.
943 Ultra High Voltage Expander is not active.
Ultra high voltage expander does not respond. The power code may be
removed.
1000 The specified module doesn’t support power compliance.
The power compliance is not available for the specified module.
1001 Illegal pulse duty
Set the pulse period and the pulse width so that the pulse duty ratio is
within the acceptable range.