User`s guide

10-44 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Error Message
Agilent FLEX Execution Error
664 Base and bias must be same polarity for I mode.
The base and bias values of the MI command must be the same
polarity.
670 Specified channel does not have ASU.
Specify the module that can be used with the ASU.
671 SSP is not available for this channel.
SSP command is available only for the CMU. Specify the slot number
that the CMU has been installed.
680 CMU correction mode must be manual.
To perform the CMU correction by using the ADJ? command, set the
CMU correction mode to manual by using the ADJ command.
681 CMU correction mode must be off.
682 Invalid standard is specified as CMU correction.
683 Frequency index is not available for CMU correction.
684 AC Voltage is 0 mV.
685 CMU correction is not complete.
690 DIO control mode must be Ultra High Current Expander / Fixture
control mode (ERMOD 4).
Set the digital IO control mode to N1265A control mode by using the
ERMOD 4 command.
691 Mismatch Ultra High Current Expander / Fixture ID.
Ultra high current expander/fixture is not connected to the Digital IO
connector. Connect the N1265A.
692 Ultra High Current Expander / Fixture control cable was disconnected.
Connection cable was removed. Turn the instrument off and connect
the cable. And then turn the instrument on again.
693 Ultra High Current Expander / Fixture is not active.
Ultra high current expander/fixture does not respond. The power switch
may be off or the power code may be removed.