Error Message
Agilent FLEX Execution Error
631 Number of search steps must be 1001 or less.
For the linear search measurement. The number of search steps
between start and stop must be 1001 or less. This means the |step| value
must be |stop-start|/1001 or more.
632 Search measurement was aborted.
Search measurement was aborted by the automatic abort function.
640 Search limits must be range/20000 or more.
For the binary search measurement. The limit value for the search
target must be range/20000 or more. where range means the
measurement range actually used for the measurement.
650 Data format must be ASCII to get time data.
The time stamp function is not available for the binary data output
format. To use the time stamp function, set the data output format to
ASCII.
655 Cannot connect/disconnect series resistor.
The series resistor status cannot be changed in the high voltage state.
Set the output voltage or the voltage compliance to the allowable
voltage for the interlock open condition or less to connect or disconnect
the series resistor.
656 Series resistor must be OFF for 1 A range.
The series resistor cannot be set to ON for the measurement channels or
the output channels that use 1 A range.
657 Series resistor cannot be used with ASU.
The series resistor is not available for the channel connected to the Atto
Sense and Switch Unit (ASU).
660 Sampling measurement was aborted.
Sampling measurement was aborted by the automatic abort function.
661 Negative hold time is only valid for I/V-t linear sampling with interval
< 2 ms.
662 Sampling interval for I/V-t log sampling must be 2 ms or longer.
663 Number of samples does not have to exceed 100001.