User`s guide
10-34 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Error Message
Agilent FLEX Execution Error
162 Incorrect command for program memory.
Specified command cannot be stored in the program memory. For the
incorrect commands, see Programming Guide.
170 Incorrect usage of internal variable.
The internal variable must be %In for integer data, or %Rn for real
data. where n is an integer, 0 to 99. Use %In for the integer type
command parameters; and use %Rn for the real type command
parameters. For the internal variables, see the VAR command of
Programming Guide.
171 Internal variable is not allowed.
The internal variables %In and %Rn are not available for the ACH,
VAR, and VAR? commands. Do not use the internal variables for the
commands.
200 Channel output switch must be ON.
To enter the specified command, set the channel output switch to ON.
201 Compliance must be set.
To change the source output mode (voltage or current), set the
compliance value.
202 Interlock circuit must be closed.
To set the output voltage or the voltage compliance to more than the
allowable voltage for the interlock open condition (high voltage state),
close the interlock circuit. If the interlock circuit is opened in the high
voltage state, outputs of all units will be set to 0 V.
203 Cannot enable channel.
The channel output switch cannot be set to ON in the high voltage state.
Set the output voltage or the voltage compliance to the allowable
voltage for the interlock open condition or less to set the switch to ON.
204 Cannot disable channel.
The channel output switch cannot be set to OFF in the high voltage
state. Set the output voltage or the voltage compliance to the allowable
voltage for the interlock open condition or less to set the switch to OFF.
Or send the CL command with no parameter to set switches of all
channels to OFF immediately.