10-30 Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1
Error Message
Agilent EasyEXPERT Operation Error
300111 EasyEXPERT database backup restoration error. Database could not be
replaced with the backup. See the log file <common data
folder>\
Agilent
\EasyEXPERT\Daemon\StaDaemon.log if needed. Source: A,
Destination: B
If Source shows the EasyEXPERT database folder name, the database
remains the same.
If Destination shows the EasyEXPERT database folder name, the
database is broken. Restore the previous database by renaming the
temporary database folder named 1.tmp1A and such.
300112 Application Library update error. Information: A
300113 The specified backup is not complete. Evidence of deletion was
detected. Specify the other backup.
300114 The specified folder is not empty. Specify an empty folder.
300115 Folder deletion was canceled. The folder is left to the following path.
Path: A
300116 EasyEXPERT database backup extraction was canceled. The extracted
files were deleted.
300117 Folder copy was canceled. The copied files were deleted.
400001 Failed to load inf.dat file. Path: A, Description: B
400002 Cannot access EasyEXPERT database. Path: A, Description: B
400003 The specified EasyEXPERT database contains the contents for another
revision. Path:
A, Description: B
400004 The specified EasyEXPERT database has the incorrect structure. Path:
A, Description: B
400005 Failed to load dir file. Path: A, Description: B
400006 Cannot access EasyEXPERT database library directory.
Path:
A, Description: B
500011 License device not found. Connect the license device for EasyEXPERT
properly. Feature: A, Error: B
500012 License device busy. Try again later. Feature: A, Error: B
500021 License for this feature is locked. Try again later. Feature: A, Error: B
500022 Unlicensed feature found. Feature: A