User`s guide
Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1 6- 5
Remote Control Interface
Introduction
Figure 6-2 Operations Supported by EasyEXPERT Remote Control Interface
Preparation for Communication
Communication between a computer and EasyEXPERT can be made as follows.
• Prepare a computer installed with VISA/SICL library. They are included in
Agilent IO Libraries Suite.
• Connect the computer to a LAN. And connect the B1500, or the computer
installed with Desktop EasyEXPERT, to the same LAN. See
Figure 6-1 for
example.
• Make a communication between the computer and EasyEXPERT using the
socket service. Only one EasyEXPERT can be connected.
If you use Agilent IO Libraries Suite, launch the Connection Expert software
and click the Add Instrument button to find and connect EasyEXPERT.
To open workspace
To set test setup
To execute measurement
To close workspace
To get test results
Exit EasyEXPERT
Start EasyEXPERT
Power ON
To change test parameters
(for Application Test)
Operations supported by
EasyEXPERT remote
control interface
(Single execution)
:WORK:OPEN
:BENC:PRES:SET:SEL
:BENC:APP:SEL
:BENC:SEL:NUMB
:BENC:SEL:STR
:BENC:SEL:RUN
:RES:FET
:WORK:CLOS