Agilent EasyEXPERT User’s Guide Vol. 2, Edition 1 10-5
Error Message
Agilent EasyEXPERT Operation Error
105033 The polarity of peak value(s) and the base value of SMU current pulse
must be the same.
105034 The start value and stop value of SMU log sweep must be non-zero and
have same polarity for primary sweep source (and synchronous sweep
source).
105035 Invalid value (A) was specified for SMU I/V Sweep stop condition.
Valid option(s): B
A is not available for the setting of the sweep abort condition. Select
one from the available choices B. If you use a formula, the calculation
result must show the available value.
105036 The hold time for SMU I/V Sweep must be between A s and B s.
105037 The delay time for SMU I/V Sweep must be between A s and B s.
105041 Invalid value (B) was specified for current monitor ranging mode of A.
Valid option(s): C
105042 Invalid value (B) was specified for current monitor range of A. Valid
option(s): C
105043 Invalid value (B) was specified for current monitor ranging rule of A.
Valid option(s): C
105044 Invalid value (B) was specified for voltage monitor ranging mode of A.
Valid option(s): C
105045 Invalid value (B) was specified for voltage monitor range of A. Valid
option(s): C
105046 Invalid value (A) was specified for the integral mode of
High-Resolution ADC. Valid option(s): B
105047 The coefficient for High Resolution ADC A integral mode must be
between B and C.
105048 Invalid value (A) was specified for the averaging mode of High-Speed
ADC. Valid option(s): B
105049 The coefficient for High Speed ADC A averaging mode must be
between B and C.
105050 Invalid value (A) was specified for the integral mode of ADC. Valid
option(s): B
105051 Invalid value (A) was specified for the sweep direction. Valid option(s):
B