User`s manual
146 Error Messages
Table C-2. Multiplexer Error Messages
Code Error Message Potential Cause(s)
-211 Trigger ignored Trigger received when scan not enabled. Trigger received after scan complete. Trigger
too fast.
-213 Init Ignored Attempting to execute an INIT command when a scan is already in progress.
-222 Data out of range Parameter value is outside valid range.
-224 Illegal parameter value Attempting to execute a command with a parameter not applicable to the command.
-240 Hardware error Command failed due to a hardware problem.
-310 System error Internal driver error. This error can result if an excessively long parameter list is entered.
1500 External trigger source already
allocated
Assigning an external trigger source to a switchbox when the trigger source has already
been assigned to another switchbox.
1510 Trigger source
non-existent
Selected trigger source is not available on this platform (e.g. some triggers are not
available on the E1300/E1301 VXI B-size mainframes).
2000 Invalid card number Addressing a module (card) in a switchbox that is not part of the switchbox.
2001 Invalid channel number Attempting to address a channel of a module in a switchbox that is not supported by the
module (e.g., channel 99 of a multiplexer module).
2006 Command not supported on this
card
Sending a command to a module (card) in a switchbox that is unsupported by the
module.
2008 Scan list not initialized Executing a scan without the INIT command.
2009 Too many channels in channel
list
Attempting to address more channels than available in the switchbox.
2010 Scan mode not allowed on this
card
The selected scanning mode is not allowed with this module or you have misspelled the
mode parameter (see SCAN:MODE command).
2011 Empty channel list No valid channels are specified in the channel_list.
2012 Invalid Channel Range Invalid channel(s) specified in SCAN <channel_list> command. Attempting to begin
scanning when no valid channel list is defined.
2600 Function not supported on this
card
Sending a command to a module (card) in a switchbox that is not supported by the
module or switchbox.
2601 Channel list required Sending a command requiring a channel_list without the channel_list.