118 Agilent E8462A Scanning Voltmeter Application Examples Chapter 4
Reset Conditions
This section describes the power-on and reset condition the E8462A switch
module is in when a “scanning voltmeter” reset occurs. The “VOLTMTR”
device driver controls both the voltmeter and any switch module configured
with it in a scanning voltmeter configuration. At power-on or following the
reset of a scanning voltmeter (
*RST command sent to the voltmeter address),
all 256 channels and the tree relays are open. All E8462A routing relay
common terminals are set to the normally closed contact. In addition, after
a
*RST command, the current scan channel list is invalidated. See the
Agilent E1326B/E1411B 5½-Digit Multimeter User's Manual for the reset
conditions of the multimeter.
Table 4-1 lists the parameters and default values for the switch module
functions following power-on or reset. These are not accessible to you via
the “VOLTMTR” driver but are provided so you know the switch module
condition after power-on or following a reset
Figure 4-1. "Scanning Voltmeter" Configuration