User`s manual

106 Agilent E8462A Relay Multiplexer Command Reference
SYST:CDES?
Return the description.
SYSTem:CPON
SYSTem:CPON <number> | ALL Sets the selected module (card) in a switchbox to
its power-on state, with the exception of the mode, interrupt line and interrupt timer
selected.
Parameters
Comments Multiplexer Module Power-on State: The power-on state is all channels (relays)
open. Note that
SYSTem:CPON ALL and *RST opens all channels of all modules in
a switchbox, while
SYSTem:CPON <number> opens the channels in only the
module (card) specified in the command. Current operating mode (as set by
FUNCtion command) will not be affected by execution of the SYSTem:CPON
<
number> or *RST commands. In addition, these commands do not affect the
DIAGnostic:INTerrupt or DIAGnostic:INTerrupt:TIMer commands.
Example Setting Card #1 Module to its Power-on State
SYST:CPON 1
Sets card #1 to power-on state.
SYSTem:CTYPe?
SYSTem:CTYPe? <number> Returns the module (card) type of a selected module
in a switchbox.
Parameters
Comments
256-Channel Multiplexer Module Model Number: The SYSTem:CTYPe?
<
number> command returns:
HEWLETT-PACKARD,E8462A,0,A.09.00
where the 0 after E8462A is the module serial number (always 0) and A.09.00 is an
example of the module revision code number.
Example Reading the Model Number of a Card #1 Module
SYST:CTYP? 1
Returns the model number.
Parameter
Name
Parameter
Type Range of Values
Default
Value
number
numeric 1 through 99 N/A
Parameter
Name
Parameter
Type Range of Values
Default
Value
number
numeric 1 through 99 N/A