Technical data
2-16
Agilent E7501A Getting Started
Performing Acceptance Test Procedures
Test 6. Verify Synchronization of the Racal 3153
Test 6. Verify Synchronization of the Racal 3153
Description This test is used to verify synchronization between all three channels of the
Racal 3153 arbitrary waveform generator.
Equipment Required • Agilent E7501A arbitrary analog signal developer
• Agilent 8563E spectrum analyzer
NOTE
A different negative detector may be used depending upon the frequency
range of the external leveling loop configuration.
Step 1. Use a high frequency 3.5 mm cable to connect the Agilent E7501A arbitrary
analog signal developer output to the input of the Agilent 8563E
spectrum analyzer.
Step 2.
Agilent E7501A Summary Agilent E7501A Details
Reset 1. From the pull down View menu,
select Reset View Settings and Windows.
Open a Signal Plan 2. From the pull down File menu,
select Open.
3. In the browser window that opens,
navigate to the folder C:\Program Files\Agilent\Measurement
and Stimulus Subsystems\Synchronization.ssp.
If you installed your software in a different path, you may need to
use the program: Start/Find/Files or Folders and enter
Synchronization.ssp in the browser window that appears.
Compile the Signal Plan 4. From the pull down Signal Plan menu,
select Compile All.
Play the Signal Plan 5. From the pull down Signal Plan menu,
select Play.