Technical data
Agilent E7501A Getting Started
2-15
Performing Acceptance Test Procedures
Test 5. Verify Hopping with Two Pulse Modulated Signals
Step 2.
Agilent E7501A Summary Agilent E7501A Details
Reset 1. From the pull down View menu,
select Reset View Settings and Windows.
Open a Signal Plan 2. From the pull down File menu,
select Open.
3. In the browser window that opens,
navigate to the folder C:\Program Files\Agilent\Measurement
and Stimulus Subsystems\FreqHopAndPulse.ssp.
If you installed your software in a different path, you may need to
use the program: Start/Find/Files or Folders and enter
FreqHopAndPulse.ssp in the browser window that appears.
Compile the Signal Plan 4. From the pull down Signal Plan menu,
select Compile All.
Play the Signal Plan 5.
From the pull down Signal Plan menu,
select Play.
To view the signal that is created, set up a
spectrum analyzer at a center frequency of
1 GHz with a 50 MHz span. Shown below are
the two signals that should be visible. Note that
the two signals may be reversed; this is
dependent on when the Racal 3153 waveform
generator triggers the hop list.