Technical data
Agilent E7501A Getting Started
2-13
Performing Acceptance Test Procedures
Test 4. Pulse Modulation Level Accuracy
Step 3.
Table 2-4 Pulse Leveled Accuracy Table
Agilent E7501A Summary Agilent E7501A Details
Frequency = 20 GHz 1. From the RF Source Control view,
select Frequency Units and set it to GHz.
2. From the RF Source Control view,
highlight the Frequency field and enter 20 using the keyboard.
Agilent 8563E Summary Agilent 8563E Details
Frequency = 20 GHz 1. Press the Frequency front panel key and enter 20 GHz.
Marker, Peak Search 2. Press the Mkr and Peak Search front panel keys.
Marker Delta, Next Peak 3. Press the Marker Delta and Next Peak front panel keys.
Record Marker value for 20 GHz
and 30 kHz PRF.
4. Enter the marker value in Table 2-4 for 20 GHz test frequency and
30 kHz PRF.
Accuracy Carrier Amplitude Pulse Amplitude–()–=
Test
Frequency
PRF Pulse Amplitude Carrier
Amplitude
Accuracy Test Limit
1 GHz 30 kHz
±
2 dB
20 GHz 30 kHz
±
2 dB