Technical data

2-12
Agilent E7501A Getting Started
Performing Acceptance Test Procedures
Test 4. Pulse Modulation Level Accuracy
Step 2.
Agilent E7501A Summary Agilent E7501A Details
Reset 1. From the pull down View menu,
select Reset View Settings and Windows.
Reference = External 2. From the RF Source Control view,
select Reference and set it to External.
Frequency = 1 GHz 3. From the RF Source Control view,
select Frequency Units and set it to GHz.
4. From the RF Source Control view,
highlight the Frequency field and enter 1 using the keyboard.
Power = 0 dBm 5. From the RF Source Control view,
highlight the Power field and enter 0 using the keyboard.
Pulse = Enabled 6. From the RF Source Control view,
select the Pulse check box.
RF On = Enabled 7. From the RF Source Control view,
select the RF Output Enable check box.
Agilent 8563E Summary Agilent 8563E Details
Preset 1. Press the green Preset front panel key.
Frequency = 1 GHz 2. Press the Frequency front panel key and enter 1 GHz.
Span = 100 kHz 3. Press the Span front panel key and enter 100 kHz.
Reference Level = +10 dBm 4. Press the Amplitude front panel key and enter +10 dBm; this sets
a reference level.
Marker, Peak Search 5. Press the Mkr and Peak Search front panel keys.
Marker Delta, Next Peak 6. Press the Marker Delta and Next Peak front panel keys.
Record Marker value for
1 GHz and 30 kHz PRF.
7. Enter the marker value in Table 2-4 for 1 GHz and 30 kHz PRF.