Technical data

Agilent E7501A Getting Started
2-9
Performing Acceptance Test Procedures
Test 3. FM Accuracy
Step 2.
Agilent E7501A Summary Agilent E7501A Details
Reset 1. From the pull down View menu,
select Reset View Settings and Windows.
Reference = External 2. From the RF Source Control view,
select Reference and set it to External.
Frequency = 10 MHz 3. From the RF Source Control view,
select Frequency Units and set it to MHz.
4. From the RF Source Control view,
highlight the Frequency field and enter 10 using the keyboard.
Power = 0 dBm 5. From the RF Source Control view,
highlight the Power field and enter 0 using the keyboard.
FM = Enabled 6. From the Stimulus Parameters view,
select the FM tab.
7. On the FM tab,
select the Value field for FM Rate and enter 200000 (200 kHz).
8. From the RF Source Control view,
select the FM check box.
RF On = Enabled 9. From the RF Source Control view,
select the RF Output Enable check box.
Agilent 8563E Summary Agilent 8563E Details
Preset 1. Press the green Preset front panel key.
Frequency = 10 MHz 2. Press the Frequency front panel key and enter 10 MHz.
Span = 100 kHz 3. Press the Span front panel key and enter 100 kHz.
Reference Level = +10 dBm 4. Press the Amplitude front panel key and enter +10 dBm; this sets
a reference level.
Marker, Peak Search 5. Press the Mkr and Peak Search front panel keys.
Marker Delta, Next Peak 6. Press the Marker Delta and Next Peak front panel keys.
Record Marker value for
10 MHz and 200 kHz FM Rate.
7. Enter the marker value in Table 2-3 for 10 MHz and 200 kHz
FM Rate.