Technical data

Agilent E7501A Getting Started
2-5
Performing Acceptance Test Procedures
Test 2. AM Accuracy
Test 2. AM Accuracy
Description This test is used to verify that AM modulation is working correctly.
The Agilent E6432A microwave synthesizer is configured for linear AM and
a spectrum analyzer is used to measure sidebands. The AM output of the
Racal 3153 arbitrary waveform generator drives the Agilent E6432A
microwave synthesizer AM input. The test is performed at carrier
frequencies of 10 MHz and 20 GHz, with AM depth of 100%, and an
AM rate of 10 kHz.
Equipment Required Agilent E7501A arbitrary analog signal developer
Agilent 8563E spectrum analyzer
Equipment Setup
NOTE
All test equipment requires a 30 minute warm-up period to ensure warranted
performance. Both the Agilent E7501A arbitrary analog signal developer
and the Agilent 8563E spectrum analyzer need to be connected to a common
10 MHz reference.
When using the Agilent E7501A arbitrary analog signal developer, ensure
that Reference is set to External; the Reference can be selected from the RF
Source Control view.
Step 1. Use a high frequency 3.5 mm cable to connect the Agilent E7501A arbitrary
analog signal developer output to the input of the Agilent 8563E
spectrum analyzer.