Technical data

2-4
Agilent E7501A Getting Started
Performing Acceptance Test Procedures
Test 1. CW Frequency and Power
Step 4.
Step 5.
Table 2-1 Maximum Power Table
Agilent E7501A Summary Agilent E7501A Details
Frequency = 10 GHz 1. From the RF Source Control view,
highlight the Frequency field and enter 10 using the keyboard.
Agilent 8563E Summary Agilent 8563E Details
Frequency = 10 GHz 1. Press the Frequency front panel key and enter 10 GHz.
Marker, Peak Search 2. Press the Mkr and Peak Search front panel keys.
Record Marker value for 10 GHz
and -40 dBm.
3. Enter the marker value in Table 2-1 for 10 GHz and -40 dBm.
Agilent E7501A Summary Agilent E7501A Details
Attenuation = 0 dB 1. From the RF Source Control view,
highlight the Attenuation field and enter 0 using the keyboard.
Agilent 8563E Summary Agilent 8563E Details
Marker, Peak Search 1. Press the Mkr and Peak Search front panel keys.
Record Marker value for 10 GHz
and 10 dBm.
2. Enter the marker value in Table 2-1 for 10 GHz and 10 dBm.
Test Frequency Selected Power Measured Power Test Limits
1 GHz 10 dBm dBm
10 GHz -40 dBm dBm
1 GHz 10 dBm dBm
10 GHz -40 dBm dBm