Technical data

Agilent E7501A Getting Started
2-3
Performing Acceptance Test Procedures
Test 1. CW Frequency and Power
Step 2.
Step 3.
Agilent E7501A Summary Agilent E7501A Details
Reset 1. From the pull down View menu,
select Reset View Settings and Windows.
Reference = External 2. From the RF Source Control view,
select Reference and set it to External.
Frequency = 1 GHz 3. From the RF Source Control view,
select Frequency Units and set it to GHz.
4. From the RF Source Control view,
highlight the Frequency field and enter 1 using the keyboard.
Power = +10 dBm 5. From the RF Source Control view,
highlight the Power field and enter 10 using the keyboard.
RF On = Enabled 6. From the RF Source Control view,
select the RF Output Enable check box.
Agilent 8563E Summary Agilent 8563E Details
Preset 1. Press the green Preset front panel key.
Frequency = 1 GHz 2. Press the Frequency front panel key and enter 1 GHz.
Span = 1 MHz 3. Press the Span front panel key and enter 1 MHz.
Reference Level = +20 dBm 4. Press the Amplitude front panel key and enter +20 dBm; this sets
a reference level.
Marker, Peak Search 5. Press the Mkr and Peak Search front panel keys.
Record Marker value for
1 GHz and 10 dBm.
6. Enter the marker value in Table 2-1 for 1 GHz and 10 dBm.
Agilent E7501A Summary Agilent E7501A Details
Auto Atten = Disabled 1. From the RF Source Control view,
unselect the Auto Atten check box so that the attenuator can be
manually controlled.
Attenuation = 50 dB 2. From the RF Source Control view,
highlight the Attenuation field and enter 50 using the keyboard.
Agilent 8563E Summary Agilent 8563E Details
Marker, Peak Search 1. Press the Mkr and Peak Search front panel keys.
Record Marker value for
1 GHz and -40 dBm.
2. Enter the marker value in Table 2-1 for 1 GHz and -40 dBm.