96 Chapter 4
Assembly Replacement and Post-Repair Procedures
Replacement of E5039B Module
Post Replacement Procedures
Following steps to update the module firmware are required after replacement of the
modules. If you don’t follow these, “revision conflict error” will be occurred when you
launch the system software.
Step 1. Turn on and log in the PC.
Step 2. Launch MS-DOS Prompt (
Start - Programs - MS-DOS Prompt) from Start menu of the
Windows 95 or 2000, and change the directory to c:\Program
Files\Agilent\E5022\Util.
Step 3. Type the command shown below. This command will download the firmware to the
module. It takes about a few minute.
Step 4. Close MS-DOS Prompt.
Operation Check
Refer to the “Perform System Initialization” on page 28 to verify the system operation.
Module Command (command_module logical address)
E5039B update39 9
[Enter]