Service manual

Chapter 4 91
Assembly Replacement and Post-Repair Procedures
Replacement of E5036A Module
1. Chapter Title
2. Chapter Title 3. Chapter Title
4. Assembly Replacement
5. Chapter Title
and Post-Repair Procedures
Replacement of E5036A Module
This part describes how to replace the E5036A module.
Tools Required
Pozidriv screwdriver
Torque Limiting Driver - 8 mm Opening 1.1 Nm (11kgfcm)
Removal
To uninstall the module from the VXI mainframe, follow below steps.
Step 1. Shut whole system down.
Step 2. Remove all cables connected to the module.
Step 3. Pull the module out from the VXI mainframe. To pull it out, loosen the screws fixing the
module to the VXI mainframe and pull the lever of the module.
Step 4. Open the faulty module and uninstall the filter matrix boards. Refer to Figure 4-1 for
details.
1. Unscrew 13 screws from the side panel.
2. Remove the side panel.
3. Remove the filter matrix boards.
NOTE Record order of these boards before removing. You should attach them on the new module
in order of the faulty module.
Replacement
To install the module into the VXI mainframe, follow below steps.
Step 1. Open the new module and install the filter matrix boards at proper position.
Step 2. Place the side panel and screw 13 screws.
Step 3. Write the faulty module’s serial number on the new module’s blank serial number tag using
a fine point permanent marker
Step 4. Insert the module into the VXI mainframe. Then, make sure the module is inserted deep
enough, and fix the module by tightening the screws on the front panel.
Step 5. Connect the cables to the module. Refer to “Cable Connection” section in this chapter.