Service manual

Chapter 2 51
Troubleshooting
Perform Measurement System Verification
1. Chapter Title
2. Troubleshooting
3. Chapter Title
4. Chapter Title
5. Chapter Title
Step 1. TAA Measurement using E5040A Spectrum Measurement Module
1. Set the fixture board on the E5029K connection board.
NOTE Do not attach or remove the verification fixture while the access lamp on the connection
board is lit. Returning to the configuration menu will turn off the access lamp and allow
you to attach or remove the fixture.
2. Connect the terminator on the ‘R’ on the fixture.
Connect the 10 dB attenuator on the ‘R+’ on the fixture.
Connect the cable between the 10 dB attenuator on the fixture board and the Ref Out on
the E5040A.
These connections are shown in Figure 2-27.
NOTE In case of installing the E4402B (opt. 300), connect the cable between the 10 dB attenuator
on the fixture board and the 10 MHz REF OUT on the E4402B instead of the Ref Out on
the E5040A.
Figure 2-27 Cable Connection of TAA Linearity Test
3. Place copies of eight verification kit setup files in the appropriate folders (if it is not
done). See Step 1-2 on page 40 for details.
4. Select
Start - Programs - Agilent Hard Disk ReadWrite Test System - Test
Environment (Run Time Version)
from the task bar of Windows 95 or 2000 in order to
start the VEE measurement program.