48 Chapter 2
Troubleshooting
Perform Measurement System Verification
Figure 2-24 Cable Connection of the E5037C/D Data Check
2. Click
Write Always to generate a write data.
3. Observe the waveform on the oscilloscope.
Figure 2-25 Data on E5037C/D
4. Read the waveform frequency. Confirm if the both frequencies (data+ and data−) are
approximately 10 MHz.
5. Click
Abort > Return > Return > Config to return to the configuration page and turn
off the output signal.
6. If the correct waveform can not be observed on the E5037C/D, replace the E5037C/D.