Service manual

Chapter 2 47
Troubleshooting
Perform Measurement System Verification
1. Chapter Title
2. Troubleshooting
3. Chapter Title
4. Chapter Title
5. Chapter Title
NOTE Remain the verification fixture on the connection pad. Otherwise an error will occur during
the measurement.
2. Click
Return > Measure > Erase & Write > Write Always to generate a write data.
Figure 2-22 Erase & Write Menu
3. Observe the waveform on the oscilloscope.
Figure 2-23 Clock Data on E5037C/D
4. Read the waveform frequency. Confirm if the frequency is approximately 100 MHz.
5. Click
Abort > Return > Return > Config to return to the configuration page and turn
off the output signal.
Step 3. Data Signal Test at E5037C/D
1. Connect the cables between the ‘Data Pos’ and ‘Data Neg’ on the E5037C/D and the
channels 1 and 2 on the oscilloscope.
These connections are shown in Figure 2-24.