Perform Measurement System Verification
1. Chapter Title
2. Troubleshooting
3. Chapter Title
4. Chapter Title
5. Chapter Title
spinstand crash.
NOTE [system installation folder] should be the folder you installed the system software. The
default installation folder is “C:\Program Files\Agilent\E5022.”
3. Select
Start - Programs - Agilent Hard Disk ReadWrite Test System - Test
Environment (Run Time Version)
from the Windows task bar to start the Test
Environment VEE program.
4. Before the program starts to initialize, the ‘E5022 VEE START’ dialog box will appear,
select the ‘Restore from file.’
Figure 2-11 Selecting at the Start Menu
5. Select the file
verification.cf0 (if you use the E5040A as a spectrum analyzer in your
system) or
verification4395a.cf0 (if you use the 4395A as a spectrum analyzer in the
your system). This setting forces the system not to initialize the spinstand.
Table 2-8 Verification Kit Setup Files and Folders
Setup File Name Folder
e5023-66510_bd.xml
[system installation folder]
\config\headamp\
e5023-66510_cal.xml
e5023-66510_ctrl.xml
e5023-66510_ic.xml
verification.cf0
[system installation folder]
\vee\
verification.cf1
verification4395a.cf0
verification4395a.cf1