Service manual

Chapter 2 33
Troubleshooting
Perform System Initialization
1. Chapter Title
2. Troubleshooting
3. Chapter Title
4. Chapter Title
5. Chapter Title
NOTE Figure 2-3 is sample of error message. If the error message return from the instrument,
VISA name and error messages appears in the PC monitor. The head of the list is the
E5023A error code regarding error message. In the case of the Figure 2-3, it is assumed
that the spinstand gives some problems.
“ASRL1::INSTR:xxxxxxxx”
VI_ERROR_RSRS_NFOUND:No
matching resource found / no more
matches
step 1 It assumes that a communication error
occurs. Turn the all instruments,
spinstand and PC on, and then perform
the system initialization.
“ASRL1::INSTR:xxxxxxxx”
hpe5022_ERROR_MEMORY:Cali
bration Data or User Data Check
Sum Failed.
step 1 Calibration data or user data in the
PMAC-PC is incorrect. Re-download
the calibration data to PMAC from the
PC.
“ASRL1::INSTR:xxxxxxxx”
hpe5022_ERROR_INTERLOCK:
Spinstand Interlocked.
step 1 Spinstand interlocked. Turn on the
E5023A, and the perform test. If the
same error message appears again, see
“Troubleshooting” chapter of spinstand
service manual.
“ASRL1::INSTR:xxxxxxxx”
hpe5022_ERROR_MOTOR_FAU
LT:Spinstand Control Error
step 1 Turn the Spinstand and PC on, and then
perform the system initialization. If the
same error message appears again, see
“Troubleshooting” chapter of spinstand
service manual.
“ASRL1::INSTR:xxxxxxxx”
hpe5022_ERROR_HARD_MISS:
Hardware Not Detected
step 1 If the media is missing on the spindle,
set it.
“ASRL1::INSTR:xxxxxxxxx”
hpe5022_ERROR_SELFTEST:Self
test Failed on the hpe5022 System
step 1 Turn the Spinstand and PC on, and then
perform the system initialization. If the
same error message appears again, see
“Troubleshooting” chapter of spinstand
service manual.
Table 2-5 List of System Initialization and Self Test Error except for VXI instruments
Error Message Trouble Isolation Procedure