Service manual
32 Chapter 2
Troubleshooting
Perform System Initialization
“GPIB0::7::INSTR:xxxxxxx”
hpe5022_ERROR_SELFTEST:Self
test Failed on hpe5022 System
step 1 Oscilloscope self test error occurred.
check and replace the Oscilloscope
"GPIB0::17::INSTR:xxxxxxxx”
VI_ERROR_RSRS_NFOUND: No
matching resources found / no more
matches.
step 1 Check the GPIB cable which is
connected with PC and 4395A or
E4402B
step 2 Check an operation of the 4395A or
E4402B, and check GPIB address.
Click on the mouse icon at the right
upper side of display and click
‘Utilities’ > ‘Remote Interface..’ by
using a mouse, then the
4395A/E4402B’s GPIB address
appears on the display. If address is not
17/18, you must enter 17/18.
step 3 Check the GPIB card within the PC
“ASRL1::INSTR:xxxxxxxx”
VI_ERROR_IO: I/O error
step 1 Turn the Spinstand and PC on once, and
perform the system initialization again.
check or replace the Serial Cable (PC
COM1 to the Spinstand).
step 2 Inspect the LED on the RS-232 to
RS-422 converter in the DMM after
system initialization. Five LED lighting
red or green is correct. If except case,
check the RS-232 cable and RS-232 to
RS-422 converter as shown in Figure
2-5.
step 3 Check the Serial I/F in the PC
“ASRL1::INSTR:xxxxxxxx” Time
out occurred
step 1 Check the serial I/F (COM 1)
configuration as shown in Figure 2-6.
(configuration of serial interface are
baud rate: 19200, Parity: None, Flow
Control: None, Stop Bits: 1, Data Size:
8 and SRQ Line: RI)
step 2 If serial I/F is no problems, it assumes
that Spinstand gives some problems.
“ASRL1::INSTR:xxxxxxxx”
VI_ERROR_FAIL_ID_QUERY:Ins
trument Identification query failed.
step 1 Spinstand hardware and DLL files are
mismatched. Re-install the DLL files to
the PC.
Table 2-5 List of System Initialization and Self Test Error except for VXI instruments
Error Message Trouble Isolation Procedure