Service manual
Chapter 2 31
Troubleshooting
Perform System Initialization
1. Chapter Title
2. Troubleshooting
3. Chapter Title
4. Chapter Title
5. Chapter Title
NOTE The Agilent VISA Assistant assigns the several instruments to the VISA name as shown in
Table 2-4. When the E5023A does not contain the E5039B/C, E5041A and/or
Oscilloscope, the VISA name of them are not assigned.
Table 2-4 VISA Name
Instrument (Interface) VISA Name
Spinstand (Serial Interface COM1) ASRL1
E8491A/B (VXI / IEEE1394) VXI0::0::INSTR
E5035B (VXI / IEEE1394) VXI0::5::INSTR
E5036A (VXI / IEEE1394) VXI0::6::INSTR
E5037C/D (VXI / IEEE1394) VXI0::7::INSTR
E5038B (VXI / IEEE1394) VXI0::8::INSTR
E5039B/C (VXI / IEEE1394) VXI0::9::INSTR
E5040A (VXI / IEEE1394) VXI0::10:INSTR
E5041A (VXI / IEEE1394) VXI0::11:INSTR
Oscilloscope (GPIB) GPIB0::7::INSTR
4395A (GPIB) GPIB0::17::INSTR
E4402B (GPIB) GPIB0::18::INSTR
Table 2-5 List of System Initialization and Self Test Error except for VXI instruments
Error Message Trouble Isolation Procedure
hpe5022_ERROR_NO_OPTION:
Option not installed
step 1 If it appears just after system software
installation, it is possible that the
oscilloscope option codeword is
incorrect. Re-install the E5023A
system software with correct codeword.
"GPIB0::7::INSTR:xxxxxxxx”
VI_ERROR_RSRS_NFOUND: No
matching resources found / no more
matches.
step 1 Check the GPIB cable which is
connected with PC and Oscilloscope
step 2 Check an operation of the Oscilloscope
(Agilent 54820A or Agilent
54845A/B), and check GPIB address.
Click on the mouse icon at the right
upper side of display and click
‘Utilities’ > ‘Remote Interface..’ by
using a mouse, then the Oscilloscope’s
GPIB address appears on the display. If
address is not 7, you must enter 7.
step 3 Check the GPIB card within the PC