Service manual

Chapter 2 29
Troubleshooting
Perform System Initialization
1. Chapter Title
2. Troubleshooting
3. Chapter Title
4. Chapter Title
5. Chapter Title
Table 2-3 List of System Initialization and Self Test Error for the VXI instruments
Error Message Trouble Isolation Procedure
“VXI0::5::INSTR:xxxxxxxx”
hpe5022_ERROR_HARD_MISS:
Hardware Not Detected
step 1 Check 36 pin I/O cable which is
connected with E5035B Head Amp
Control and Head Amp.
step 2 Check four terminal cable which is
connected with the E5035B Spinstand
Control and HLM Control.
step 3 Check the E5035B
hpe5022_ERROR_HARD_MISS:
Hardware Not Detected
step 1 In case of only error message appears,
but VXI instrument number (VISA
name) does not appear, check four
terminal cable which is connected with
the E5035B Spinstand Control and
Spinstand Piezo Control.
step 2 Replace the E5035B
“VXI0::5::INSTR:xxxxxxxx”
hpe5022_ERROR_MEMORY:
Calibration Data or User Data
Check Sum Failed
step 1 It is assumed that the calibration data in
the Head Amp is lost. replace the Head
Amp.
step 2 Check 36 pin I/O cable which is
connected with the E5035B Head Amp
Control and Head Amp Control.
step 3 Replace the E5035B.
“VXI0::x::INSTR:xxxxxxxx”
VI_ERROR_RSRS_NFOUND:No
matching resource found / no more
matches. (where ‘x’ is 5,6,7,8 or 9)
step 1 Check the E5035B, E5036A,
E5037C/D, E5038B or E5039B/C.
“VXI0::x::INSTR:xxxxxxxx”
VI_ERROR_PRES_NFOUND:No
matching resource found / no more
matches (all VXI modules error
appears)
step 1 Check IEEE-1394 cable which is
connected with PC IEEE-1394 port and
the E8491A/B IEEE1394 Port.
step 2 Check the E8491A/B as shown in
Figure 2-4.
step 3 Check the E8401A (VXI Main Frame)
“VXI0::x::INSTR:xxxxxxxx”
hpe5022_ERROR_REVISION:
Sub DLLs and/or module firmware
revision conflict. (where ‘x’ is
5,6,7,8 or 9)
step 1 E5023A system software (DLL) and
module firmware conflict. update VXI
module firmware which is included in
the installer of E5023A system
software.