Service manual
Appendix D 175
Error Messages
Error Messages on PC Monitor Display
1. Chapter Title
2. Chapter Title 3. Chapter Title
D. Error Messages
5. Chapter Title
hpe5022_ERROR_FILE_ACCESS Data log file in the auto
alignment operation can not
be created.
hpe5022_ERROR_FUNCTION_NOT_LOADED Unable to load the DLL file.
Perform the E5023A
installation software or
check if the file is installed
in the PC. The location of
DLL file is described in
Chapter 2 of the
programming manual.
hpe5022_ERROR_HARD Detected a hardware error
in the system.
hpe5022_ERROR_HARD_HAMP Detected a hardware in the
head amplifier board.
hpe5022_ERROR_HARD_MISS The media or HGA cassette
is missing or the selected
hardware is not installed.
Check if they are placed
properly.
hpe5022_ERROR_INACCURATE_BER Too many Bit Error Rate
happens and measurement
result is not correct. Need to
optimize the measurement.
hpe5022_ERROR_INTERNAL Unexpected internal error
occurred.
hpe5022_ERROR_INTERLOCK The spinstand is out of air
or vacuum or emergency
switch has been turned on.
Check if air or vacuum is
supplied to the spinstand or
reset the emergency switch.
hpe5022_ERROR_INV_BAS_HNDL Detected the invalid base
segment handle parameter
in UDS. Check your
program.
hpe5022_ERROR_INV_BAS_ID Detected the invalid base
segment ID parameter in
UDS. Check your program.
Error Messages on PC Monitor Display
Error Code Description