Software Recovery
Installing Agilent I/O Library
1. Chapter Title
2. Chapter Title 3. Chapter Title
4. Chapter Title
5. Software Recovery
• VXI0:6:INSTR (E5036A)
• VXI0:7:INSTR (E5037C/D)
• VXI0:8:INSTR (E5038B)
• VXI0:9:INSTR (E5039B/C)
• VXI0:10:INSTR (E5040A)
• VXI0:11:INSTR (E5041A)
4. Confirm if the GPIB control properly works if the 54845A Oscilloscope is connected.
a. Select
GPIB0::7::INSTR in the left side of the screen.
b. Select
Formatted I/O tab.
c. Check IEEE 488.2.
d. Click
*RST.
e. Click
*IDN? to confirm if the serial number data of 54845A is returned.
5. Confirm if the GPIB control properly works if the 4395A Spectrum Analyzer is
connected.
a. Select
GPIB0::17::INSTR in the left side of the screen.
b. Select
Formatted I/O tab.
c. Check IEEE 488.2.
d. Click
*RST.
e. Click
*IDN? to confirm if the serial number data of 4395A is returned.
6. Confirm if the GPIB control properly works if the E4402B Spectrum Analyzer is
connected.
a. Select
GPIB0::18::INSTR in the left side of the screen.
b. Select
Formatted I/O tab.
c. Check IEEE 488.2.
d. Click
*RST.
e. Click
*IDN? to confirm if the serial number data of E4402B is returned.