Service manual

132 Chapter 4
Assembly Replacement and Post-Repair Procedures
How to Check the Operation After Replacement
It is able to confirm the E5023A system configuration, such as instrument serial number,
software revisions and driver revisions.
Software Revision and Hardware Information
The revision of the E5023A system software, the serial number of modules and instrument
are displayed using the E5023A VEE measurement program. It explains how to display the
software revisions and hardware information using it
1. Select
Start - Programs - Agilent Hard Disk ReadWrite Test System - Test
Environment (Run Time Version)
from Start menu of Windows 2000 to launch the
program.
2. Before the program starts to initialize, the ‘VEE Start’ will appear. Then select ‘Starts
with the default setting’. Initialization takes about 60 seconds.
3. After initialization, the ‘cassette Parameter Setup’ menu will pop up. Click the
Close
button.
4. Click the
About Agilent E5022/E5023 button to display the current software versions
and hardware informations as shown in Figure 4-22.
Figure 4-22 About Agilent E5022/E5023A Display
NOTE Since the E5036A (VXI0::6::INSTR) has no internal DSP chip, there is no firmware to be
update. For the spinstand (ASRL1::INSTR), module revision and serial number are not
returned.
*6.If exchanged head cassette is not same as existed one, it is necessary to re-enter the cas-
sette parameter.
*7.After replacing an assembly of spinstand system, you may additionally need to perform
re-adjustment. See spinstand service manual for details.