Service manual

Chapter 4 131
Assembly Replacement and Post-Repair Procedures
How to Check the Operation After Replacement
1. Chapter Title
2. Chapter Title 3. Chapter Title
4. Assembly Replacement
5. Chapter Title
and Post-Repair Procedures
How to Check the Operation After Replacement
After replacement of the instrument, it is necessary to perform the software updating,
re-configuration and re-adjustment as shown in Table 3-11. And, perform the System
Check function or user application program again to verify the E5023A operation.
Table 4-15 Requirements after replacement of instruments (---: not required)
exchange instruments
s/w update re-config.
PC
x
*1
x
*2
GPIB I/F card in the PC
x
*3
x
*2
IEEE 1394 I/F card in the PC
x
*3
x
*2
E8401A (VXI Main Frame) --- ---
E8491B (IEEE 1394 PC Link to VXI)
x
*3
x
*2
E5035B (Spinstand Interface Module)
x
*4
---
E5036A (Filter Matrix Module) --- ---
E5037C/D (Data Generator Module)
x
*4
---
E5038B (Parametric Measurement
Module)
x
*4
---
E5039B/C (Bit Error Test Module)
x
*4
---
E5041A (Dual Counter Module) --- ---
E5040A (Spectrum Analyzer Module)
x
*4
---
54845A/B (Oscilloscope) --- ---
E4402B (Spectrum Analyzer) --- ---
Head Amplifier
x
*5
---
HGA Cassette
x
*5
x
*6
Spinstand Go to troubleshooting chapter in
spinstand service manual
*7
*1.All software (Agilent I/O library, Agilent VEE, E5022/E5023 system software, Acrobat
Reader).
*2.The I/O card must be re-configured using the Agilent I/O config.
*3.It is necessary to update the Agilent I/O library depending on the situation, for example
exchanged the interface card if it is different from the present one.
*4.If system DLLs and module firmware revision conflict, it is necessary to update a
instrument firmware or to install the E5023A system software again.
*5.If ahead amplifier board or cassette is changed, the update of the configuration files is
required.