Service manual
Chapter 4 109
Assembly Replacement and Post-Repair Procedures
Cable Connection
1. Chapter Title
2. Chapter Title 3. Chapter Title
4. Assembly Replacement
5. Chapter Title
and Post-Repair Procedures
3 FFC Cable E5029K Connection Board
“Connection Board I/F”
E5029K Connection Board
“Cable”
E5029-61603
4 SMA(m)-SMA(m) Cable E5043A “OUT D+” E5029K “W+”
E5043-61641
*1
5 SMA(m)-SMA(m) Cable E5043A “OUT D−” E5029K “W−”
E5043-61642
*1
6 SMA(m)-SMA(m) Cable E5036A “Sig In” Cable “7” E5023-61607
7 SMA(m)-SMA(m) Cable Cable “6” E5029K “Read Sig Out” E5023-61608
8 50P Halfpitch Flat Cable E5029K Onstage Buffer Unit E5043A Head Amplifier
Control Unit (Main Unit)
E5043-61613
9 SMA(m)-SMA(m) Cable E5039B/C “Data Out Pos” E5043A “Input 2 Data+” E5039-61606
10 SMA(m)-SMA(m) Cable E5039B/C “Data Out Neg” E5043A “Input 2 Data−” E5039-61607
11 SMA(m)-SMA(m) Cable E5037C “Clk Out Pos” E5043A “Input Clock+” E5023-61601
12 SMA(m)-SMA(m) Cable E5037C “Clk Out Neg” E5043A “Input Clock−” E5023-61602
13 SMA(m)-SMA(m) Cable E5037C “Data Out Pos” E5043A “Input 1 Data+” E5023-61603
14 SMA(m)-SMA(m) Cable E5037C “Data Out Neg” E5043A “Input 1 Data−” E5023-61604
15 MDR-MDR Cable E5035B “Head Amp Ctrl” E5043A “Head Amp Ctrl” E5023-61605
32 SMA(m)-SMA(m) Cable E5029K Connection Board
“W+”
E5029K Onstage Buffer Unit
“W+”
E5029-61604
33 SMA(m)-SMA(m) Cable E5029K Connection Board
“W−”
E5029K Onstage Buffer Unit
“W−”
E5029-61605
*1.E5043-61611 and E5043-61612 were replaced with these E5043-61641 and E5043-61642 respectively in Decem-
ber 2002.
Table 4-1 Cable Connection (Head Amplifier, HLM) (E5023A-415, 1.5 Gbps)
No. Description Connection Part Number