Service manual

104 Chapter 4
Assembly Replacement and Post-Repair Procedures
PC (System Controller) Replacement
PC (System Controller) Replacement
Tools Required
Pozidriv screwdriver
Removal
Step 1. Remove all cables (GPIB cable, IEEE-1394 cable, and serial cable) from the PC.
Step 2. Remove GPIB card from the PC.
Step 3. Remove IEEE-1394 card from the PC.
Replacement
Step 1. Install the Agilent E5023A software including Agilent I/O Library into the PC. See
Chapter 5, “Software Recovery,” on page 133 for the installation procedure.
NOTE You must install Agilent I/O Library into the PC before installing the GPIB and IEEE-1394
cards.
Step 2. Install the GPIB card which was used for the old PC into a PCI slot of the new PC.
Step 3. Install the IEEE-1394 card which was used for the old PC into a PCI slot of the new PC.
NOTE Generally, you can install a PCI card into any slot in the PC. However, the card position
might cause the problem in case of combination of GP-IB and IEEE-1394 cards.
From the rear view, the IEEE-1394 card must be installed in the most upper (or left) slot
and the GP-IB card must be installed in the second slot from the top (or left).
There is a cable on the IEEE-1394 PCI card to supply power from the connector on the
FDD or the CD-ROM drives. The cable has 4-pin-female connector on one side and has a
4-pin-female and another 4-small-pin-female connectors on the other side in order to split
the power. The 4-small-pin-female is connect with IEEE-1394 card. When you replace the
IEEE-1394 card, you have to remove the cable from the defective PC, too. Then install
them into a new PC.
Step 4. Connect GPIB cable, IEEE-1394 cable, and serial cable to the PC. See “PC (System
Controller) Replacement” on page 104 for how to connect those cables.