Agilent E5023A Hard Disk Read/Write Test System Service Manual Fourth Edition Part No.
Notices The information contained in this document is subject to change without notice. This document contains proprietary information that is protected by copyright.All rights are reserved. No part of this document may be photocopied, reproduced, or translated to another language without the prior written consent of the Agilent Technologies. Agilent Technologies Japan, Ltd. Component Test PGU 1-3-2, Murotani, Nishi-Ku, Kobe-shi, Hyogo, 651-2241 Japan Agilent Technologies Japan, Ltd.
Do not operate the instrument in the presence of flammable gasses or fumes. Operation of any electrical instrument in such an environment constitutes a definite safety hazard. • Keep Away From Live Circuits Operating personnel must not remove instrument covers. Component replacement and internal adjustments must be made by qualified maintenance personnel. Do not replace components with the power cable connected. Under certain conditions, dangerous voltages may exist even with the power cable removed.
condition or the like, which, if not correctly performed or adhered to, could result in injury or death to personnel. CAUTION This Caution sign denotes a hazard. It calls attention to a procedure, practice, condition or the like, which, if not correctly performed or adhered to, could result in damage to or destruction of part or all of the product. NOTE Note denotes important information. It calls attention to a procedure, practice, condition or the like, which is essential to highlight.
Assistance Product maintenance agreements and other customer assistance agreements are available for Agilent Technologies products. For any assistance, contact your nearest Agilent Technologies Sales and Service Office. Addresses are provided at the back of this manual. Typeface Conventions Bold Boldface type is used when a term is defined. For example: icons are symbols. Italic Italic type is used for emphasis and for titles of manuals and other publications.
Contents 1. Overview Manual Overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12 About this Manual. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13 Name of Each Part . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14 System Overview . . .
Contents Post Replacement Procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 89 Operation Check . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 90 Replacement of E5036A Module . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 91 Tools Required . . . . . . . . . . . . . . . . . . . .
Contents Connecting the 3 GHz Spectrum Analyzer (Option) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 128 How to Check the Operation After Replacement . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 131 Software Revision and Hardware Information . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 132 5. Software Recovery Installing Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
Contents 10
1. Overview 2. Chapter Title 3. Chapter Title 4. Chapter Title Overview 1 5.
Overview Manual Overview Manual Overview The Agilent E5023A has the following manuals: Site Preparation and Installation Manual This manual explains the necessary procedures for site preparation and installation of the system. Information on how to unpack the equipment, how to set the system, proper cable connections, air pressure and vacuum settings are all explained in this manual.
Overview Manual Overview This manual provides information on troubleshooting and servicing the Agilent E5023A Hard Disk Read/Write Test System. Service information on spinstand is not included in this manual. 1. Overview About this Manual Chapter 1, “Overview,” on page 11 This chapter explains the manual structure, product overview and name of each part of the system. This chapter describes over all troubleshooting flow.
Overview Name of Each Part Name of Each Part The name of each part of Agilent E5023A are listed below. The name of each measuring instrument are listed in Figure 1-1. If rack option is not ordered, the rack will not be provided. The name of each measuring instrument installed in the VXI mainframe are as follows.
Overview Name of Each Part 1. Overview Figure 1-1 Name of each Measuring Instrument 2. Chapter Title 3. Chapter Title 4. Chapter Title 5.
Overview Name of Each Part Figure 1-2 Spinstand (E5013A) Overview 16 Chapter 1
Overview System Overview 1. Overview System Overview This section explains the structure of the system. The system block diagram is shown in Figure 1-3. 2. Chapter Title 3. Chapter Title 4. Chapter Title 5.
Overview System Overview Figure 1-3 System Block Diagram (Read/Write Signal Flow) 18 Chapter 1
Overview System Overview 1. Overview Measuring Instrument System The measuring instrument system of the Agilent E5023A has three major parts. • • • Agilent E5037C/D Data Generator Module • Agilent E5039B/C Bit Error Rate Test Module Signal Distribution Part • Agilent E5043A/C Head Amplifier Control Unit • Agilent E5029K Onstage Buffer Unit • Agilent E5036A Filter Matrix Module 2.
Overview System Overview Control (PC) System The commands for each measuring instrument and spinstand all come from the PC. The commands being sent by the PC to each module of the VXI mainframe passes through the IEEE-1394 interface.The triggers synchronizing with the index signal are generated from Agilent E5035B Spinstand Interface Module to all modules and instrument.
Overview System Overview 1. Overview measurement library controls this. Agilent VEE Agilent E5023A VEE is a powerful visual programming language, which allows you to control the instrument, make quick measurements and display the system settings. This software controls the system by calling the Agilent E5023A measurement library. Agilent E5023A Test Environment (VEE-based graphical operating program) 2.
Overview System Overview 22 Chapter 1
1. Chapter Title 2. Troubleshooting 3. Chapter Title 4. Chapter Title Troubleshooting 2 5.
Troubleshooting Start Here Start Here The information in this chapter helps you identify the portion of the system that is at fault. The troubleshooting strategy of this manual is based on a verification (rather than symptomatic). This chapter describes troubleshooting procedure step by step, and you can find next necessary action by following the troubleshooting steps sequentially.
Figure 2-1 Primary Trouble Isolation Flowchart 2. Troubleshooting 3. Chapter Title 4. Chapter Title 5. Chapter Title 25 Chapter 2 1.
Troubleshooting Perform Power-on Self Test Perform Power-on Self Test The actions in the primary trouble isolation are done in the power-on sequence. Turn the system power on and inspect the status of the instruments (status LED on the front panel and error message in the LCD of the spinstand) as reference to the Table 2-1. If you find a symptom of the problems, it is necessary to isolate a defective instrument as reference to trouble isolation flow.
Table 2-1 Inspect The Power On Sequence Asserted Instrument Action (check or replace instrument) 54845A/B (Oscilloscope) No Display (normally Display on) 54845A/B E4402B (Spectrum Analyzer) No Display (normally Display on) E4402B Spinstand See spinstand service manual. 2. Troubleshooting Figure 2-2 Symptom of the Problem 1. Chapter Title Troubleshooting Perform Power-on Self Test VXI Main Frame and Instruments Front Panel 3. Chapter Title 4. Chapter Title 5.
Troubleshooting Perform System Initialization Perform System Initialization In this section, inspect initialization of the system. Launch the pre-installed Agilent E5023A Measurement Software and confirm if the system can initialize itself properly in the initialize routine of the program. This achieves by calling initialize function at the beginning of the program. Start the program and watch for the following events in this order. 1.
Table 2-3 List of System Initialization and Self Test Error for the VXI instruments Check 36 pin I/O cable which is connected with E5035B Head Amp Control and Head Amp. step 2 Check four terminal cable which is connected with the E5035B Spinstand Control and HLM Control. step 3 Check the E5035B step 1 In case of only error message appears, but VXI instrument number (VISA name) does not appear, check four terminal cable which is connected with the E5035B Spinstand Control and Spinstand Piezo Control.
Troubleshooting Perform System Initialization Table 2-3 List of System Initialization and Self Test Error for the VXI instruments Error Message Trouble Isolation Procedure “VXI0::x::INSTR:xxxxxxxxx” hpe5022_ERROR_SELFTEST:Self test Failed on hpe5022 System (where ‘x’ is 5,6,7,8,or 9) step1 30 VXI instruments self test error occurred. check and replace a defective VXI instrument.
NOTE The Agilent VISA Assistant assigns the several instruments to the VISA name as shown in Table 2-4. When the E5023A does not contain the E5039B/C, E5041A and/or Oscilloscope, the VISA name of them are not assigned.
Troubleshooting Perform System Initialization Table 2-5 List of System Initialization and Self Test Error except for VXI instruments Error Message Trouble Isolation Procedure “GPIB0::7::INSTR:xxxxxxx” hpe5022_ERROR_SELFTEST:Self test Failed on hpe5022 System step 1 Oscilloscope self test error occurred. check and replace the Oscilloscope "GPIB0::17::INSTR:xxxxxxxx” VI_ERROR_RSRS_NFOUND: No matching resources found / no more matches.
Table 2-5 Trouble Isolation Procedure “ASRL1::INSTR:xxxxxxxx” VI_ERROR_RSRS_NFOUND:No matching resource found / no more matches step 1 It assumes that a communication error occurs. Turn the all instruments, spinstand and PC on, and then perform the system initialization. “ASRL1::INSTR:xxxxxxxx” hpe5022_ERROR_MEMORY:Cali bration Data or User Data Check Sum Failed. step 1 Calibration data or user data in the PMAC-PC is incorrect. Re-download the calibration data to PMAC from the PC.
Troubleshooting Perform System Initialization Figure 2-3 Sample of Error Dialog Box NOTE When the E8401A VXI Main Frame or PC is turned on, the VXI Resource Manager executes automatically. If VXI Resource Manager fails, a warning message appears such as the Figure 2-4. See a detail of the error messages and make sure the VXI module’s status using the Agilent I/O Libraries. If you find a defective module, replace it.
Select Start - Programs - Agilent I-O Libraries - I-O Config from Windows Start menu to launch the program. Double Click on the label of “COM1 ASRL1” in the configured interface. Then the ‘RS-232 / Serial Configuration’ menu will pop up as shown in Figure 2-6. Figure 2-6 1. Chapter Title Troubleshooting Perform System Initialization RS-232 / Serial Configuration Menu 2. Troubleshooting 3. Chapter Title 4. Chapter Title 5.
Troubleshooting Perform Measurement System Verification Perform Measurement System Verification The measurement system verification verifies the E5023A’s basic measurement functions through the write and read signal measurements using the verification fixture. This verification does not verify the head amplifier.
Table 2-6 Contents of P/N E5023-65600 Verification Kit (commercial HA-type) 1.
Troubleshooting Perform Measurement System Verification Figure 2-7 and Figure 2-8 show the top and bottom views of P/N E5023-65520 and P/N E5023-65509. Figure 2-7 P/N E5023-65520 Verification fixture (commercial head amplifier type) Figure 2-8 P/N E5023-65509 Verification fixture (universal head amplifier type) Write Portion Check Description The measurement system verification verifies if the write data generates. The write data is observed by the oscilloscope.
Figure 2-9 Write Portion Check Signal Flow 1. Chapter Title Troubleshooting Perform Measurement System Verification 2. Troubleshooting 3. Chapter Title 4. Chapter Title 5.
Troubleshooting Perform Measurement System Verification Step 1. Write Signal Test at Connection Board 1. Connect the cables between the ‘W+’ and ‘W-’ on the fixture board and the channels 1 and 2 on the oscilloscope. NOTE Do not attach or remove the verification fixture while the access lamp on the connection board is lit. Returning to the configuration menu will turn off the access lamp and allow you to attach or remove the fixture. These connections are shown in Figure 2-10.
spinstand crash. Table 2-8 Verification Kit Setup Files and Folders Setup File Name 1. Chapter Title Troubleshooting Perform Measurement System Verification Folder e5023-66510_bd.xml e5023-66510_cal.xml 2. Troubleshooting e5023-66510_ctrl.xml [system installation folder] \config\headamp\ e5023-66510_ic.xml verification.cf0 verification.cf1 verification4395a.cf0 [system installation folder] \vee\ verification4395a.cf1 NOTE 3.
Troubleshooting Perform Measurement System Verification Figure 2-12 Selecting the Setup File (when you use the E5040A in the system) 6. Initialization takes about 30 seconds for the program to initialize. 7. After initialization, the ‘HGA Parameter Setup’ menu will pop up. Click Close. Figure 2-13 Closing the HGA Parameter Setup 8. Click Configuration. Then click Setup to open the Setup menu.
Figure 2-14 Main Menu 1. Chapter Title Troubleshooting Perform Measurement System Verification 2. Troubleshooting 9. Set the channel bit rate to 100 Mbps and the data pattern to LF. Figure 2-15 Setup Menu 3. Chapter Title 4. Chapter Title 10. Click Pattern Config to open the Data Pattern Menu. Set the LF at 5T and confirm if the channel bit rate set at 100 Mbps. 5.
Troubleshooting Perform Measurement System Verification Figure 2-16 Data Pattern Menu 11. Click Return three times to return the Main Menu. Click Measure. Figure 2-17 Main Menu 12. Click Erase & Write to open the Erase and Write Menu.
Figure 2-18 Main Menu 1. Chapter Title Troubleshooting Perform Measurement System Verification 2. Troubleshooting 13. Click Write Always to generate a write data. Figure 2-19 Erase & Write Menu 3. Chapter Title 4. Chapter Title 14. Observe the waveform on the oscilloscope. 5.
Troubleshooting Perform Measurement System Verification Figure 2-20 Write Data 15. Read the waveform frequency. Confirm if the both mean frequencies (write+ and write−) are 10 MHz ± 100 Hz. 16. Click Abort to stop the data pattern. 17. Click Return > Return > Config to return to the configuration page and turn off the output signal. 18. If the waveform is not observed, check the cable connection. Especially, clean the E5029K connection board and the connection pads on it. Step 2.
NOTE Remain the verification fixture on the connection pad. Otherwise an error will occur during the measurement. 1. Chapter Title Troubleshooting Perform Measurement System Verification 2. Click Return > Measure > Erase & Write > Write Always to generate a write data. Figure 2-22 Erase & Write Menu 2. Troubleshooting 3. Chapter Title 3. Observe the waveform on the oscilloscope. Figure 2-23 Clock Data on E5037C/D 4. Chapter Title 4. Read the waveform frequency.
Troubleshooting Perform Measurement System Verification Figure 2-24 Cable Connection of the E5037C/D Data Check 2. Click Write Always to generate a write data. 3. Observe the waveform on the oscilloscope. Figure 2-25 Data on E5037C/D 4. Read the waveform frequency. Confirm if the both frequencies (data+ and data−) are approximately 10 MHz. 5. Click Abort > Return > Return > Config to return to the configuration page and turn off the output signal. 6.
Otherwise, replace the E5043A/C. Table 2-9 Faulty Assembly Isolation Measurement Results 1.
Troubleshooting Perform Measurement System Verification Figure 2-26 Read Portion Check Signal Flow 50 Chapter 2
Step 1. TAA Measurement using E5040A Spectrum Measurement Module 1. Set the fixture board on the E5029K connection board. NOTE 1. Chapter Title Troubleshooting Perform Measurement System Verification Do not attach or remove the verification fixture while the access lamp on the connection board is lit. Returning to the configuration menu will turn off the access lamp and allow you to attach or remove the fixture. 2. Connect the terminator on the ‘R−’ on the fixture.
Troubleshooting Perform Measurement System Verification 5. Before the program starts to initialize, the ‘E5022 VEE START’ will appear, select the ‘Restore from file’. Figure 2-28 Selecting at the Start Menu 6. Select the file verification.cf0 (if you use the E5040A as a spectrum analyzer in your system) or verification4395a.cf0 (if you use the 4395A as a spectrum analyzer in the your system). This setting forces the system not to initialize the spinstand.
Figure 2-30 Closing the HGA Parameter Setup 1. Chapter Title Troubleshooting Perform Measurement System Verification 2. Troubleshooting 9. Click Configure - Setup to open the Setup menu. Figure 2-31 Main Menu 3. Chapter Title 4. Chapter Title 5.
Troubleshooting Perform Measurement System Verification Figure 2-32 Configuration Menu 10. Set the channel bit rate at 100 Mbps and the data pattern at LF. Figure 2-33 Setup Menu 11. Click Input Range to open Input Range Menu. Set the Measurement Mode at Manual Mode, the input range at 300,000µ (300m) V, and the PW reference 33.000ns. Then click Return to return the setup menu.
Figure 2-34 Input Range Menu 1. Chapter Title Troubleshooting Perform Measurement System Verification 2. Troubleshooting 12. Click Filter Config to open the Filter Menu. Select the low pass filter of the highest cut-off frequency in the installed filters. Then click Return. Filter Menu 3. Chapter Title Figure 2-35 4. Chapter Title 13. Click Pattern Config to open the Data Pattern Menu. Set the LF at 5T and confirm if the channel bit rate set at 100 Mbps. 5.
Troubleshooting Perform Measurement System Verification Figure 2-36 Data Pattern Menu 14. Click Return three times to return the Main Menu, then click Measure. Figure 2-37 Main Menu 15. Click Measure Spectrum to open the Spectrum Menu.
Figure 2-38 Measure Menu 1. Chapter Title Troubleshooting Perform Measurement System Verification 2. Troubleshooting 16. Set the parameters as shown in Figure 2-39. Figure 2-39 Spectrum Measurement Menu 3. Chapter Title 4. Chapter Title 17. Click Measure to make a spectrum measurement. Although some error message may be displayed because the spinstand is not initialized, ignore it 18. Click the marker in the graph, then move it to the peak level. 19. Read its frequency (f) and level (x).
Troubleshooting Perform Measurement System Verification Step 2. TAA Measurement using E5038B Parametric Measurement Module 1. Click Return to return the Measure Menu. 2. Click TAA to open the TAA Menu. Figure 2-40 Measure Menu 3. Set the parameters as shown in Figure 2-41. Figure 2-41 TAA Menu 4. Click Measure to start a TAA measurement. Although some error message may be displayed because the spinstand is not initialized, ignore it Confirm the measured TAA is approximately same as the calculated TAA.
NOTE In case of installing the E4402B (opt. 300), if the spectrum measurement result is not good, use the Ref Out on the E5040A and check the E5040A with “Step 1. TAA Measurement using E5040A Spectrum Measurement Module” on page 51 and “Step 2. TAA Measurement using E5038B Parametric Measurement Module” on page 58. If the E5040A’s measurement result is good, a failure in the E4402B is suspected. 1.
Troubleshooting Perform Measurements Perform Measurements This section helps you to isolate a problem in the system through actual device measurements. In brief, the Hard Disk Read/Write Test System consists of a data generation instruments, measuring instruments and the spinstand mechanical part as shown in Figure 2-42 and Figure 2-43 (Agilent E5023A does not include the spinstand). These figures are useful to isolate a defective assembly in the system.
Figure 2-42 Agilent E5023A Write/Read Data Flow 1. Chapter Title Troubleshooting Perform Measurements 2. Troubleshooting 3. Chapter Title 4. Chapter Title 5.
Troubleshooting Perform Measurements Figure 2-43 Agilent E5023A Control/Trigger/Gate Line 62 Chapter 2
Table 2-11 Trouble Isolation Hints Auto Configuration “Auto Configuration Error” occurs or TAA measurement data and/or PW measurement data is unacceptable. If a parameter of write and sense current data is unsuitable, it is assumed that the head amplifier, E5035B or E5038B has some problem. Read/Write Offset Measurement “No Peak” error message occurs during the measurement or track profile data is unacceptable.
Troubleshooting Perform Measurements Table 2-11 Trouble Isolation Hints Spectral SNR, Measure Spectrum with E5040A Noise Level is high Verify the noise level at the ‘Signal In’ connector on the E5040A. If it is out of limit, replace the E5040A. If it is acceptable, specify the noise source in the E5023A system.
Table 2-11 Trouble Isolation Hints Measurement data is unacceptable If popcorn noise data measured by the E5041A is wrong, replace the E5041A. If the E5041A replacement does not solve the problem, check unwanted noise source generated within the measurement system and remove it. All measurement Items Measurement data is not returned If the status LED on the VXI module lighting is amber, replace the module.
Troubleshooting Perform Measurements 66 Chapter 2
1. Chapter Title 2. Chapter Title 3. Replaceable Parts 4. Chapter Title Replaceable Parts 3 5.
Replaceable Parts Introduction Introduction In this chapter, the replaceable parts used for on-site repair are listed. How to order the parts is also described. This chapter lists the replaceable parts for followings. • Replaceable VXI modules and equipment • Cables and connectors In this chapter, replaceable parts for the following instruments are not described, since these instruments are repaired at service center.
Ordering Information 1. Chapter Title Replaceable Parts Ordering Information To order a part listed in the replaceable parts table, quote the Agilent part number (with a check digit), indicate the quantity required, and address the order to the nearest Agilent office. The check digit will ensure accurate and timely processing of the order. Direct Mail Order System Within the USA, Agilent can supply parts through a direct mail order system. Advantages of using this system are: 2.
Replaceable Parts Replaceable Parts Replaceable Parts Replaceable VXI Modules and Equipment Figure 3-1 Replaceable VXI Modules and Equipment Table 3-1 VXI Modules Part Number CD Qty.
Table 3-1 VXI Modules Part Number 3 5 7 8 Description E5037-65004 1 E5037C 1.5 Gbps Data Generator Module E5037-69004*1 1 E5037C 1.5 Gbps Data Generator Module (exchange) E5037-65005 1 E5037D 2.6 Gbps Data Generator Module E5037-69005*1 1 E5037D 2.
Replaceable Parts Replaceable Parts Table 3-2 Other Equipment Part Number 14 15 16 CD Qty.
Table 3-3 Filter Boards for the E5036A Filter Matrix Module Part Number CD Qty. Description 10 E5036-66578 0 1 15 MHz Low Pass Filter Board 11 E5036-66579 1 1 30 MHz Low Pass Filter Board 12 E5036-66598 4 1 Blank Filter Board (for user design) 13 E5036-66599 5 1 50 ohm Thru Filter Board 1. Chapter Title Replaceable Parts Replaceable Parts 2. Chapter Title 3. Replaceable Parts 4. Chapter Title 5.
Replaceable Parts Replaceable Parts Cables and Connectors of E5023A Option 415 (1.5 Gbps)(without opt. 300) Refer to the “Cable Connection” on page 106 for more details.
Figure 3-2 Cable Connection of E5023A Option 415 (1.5 Gbps) (without opt. 300) 1. Chapter Title Replaceable Parts Replaceable Parts 2. Chapter Title 3. Replaceable Parts 4. Chapter Title 5.
Replaceable Parts Replaceable Parts Table 3-4 Cable Connection of E5023A Option 415 (1.5 Gbps) (without opt. 300) No.
Table 3-4 Cable Connection of E5023A Option 415 (1.5 Gbps) (without opt.
Replaceable Parts Replaceable Parts Cables and Connectors of E5023A Option 426 (2.6 Gbps) (without opt. 300) Refer to the “Cable Connection” on page 106 for more details.
Figure 3-3 Cable Connection of E5023A Option 426 (2.6 Gbps) (without opt. 300) 1. Chapter Title Replaceable Parts Replaceable Parts 2. Chapter Title 3. Replaceable Parts 4. Chapter Title 5.
Replaceable Parts Replaceable Parts Table 3-5 Cable Connection of E5023A Opt. 426 (2.6 Gbps) (without opt. 300) No.
Table 3-5 Cable Connection of E5023A Opt. 426 (2.6 Gbps) (without opt. 300) No. Description Connection 29 BNC(m)-BNC(m) Cable E5035B “Index In” Spinstand Index E5023-61609 30 SMA(m)-SMB(m) Cable E5035B “Piezo Ctrl” PZT Ctrl Analog E5035-61601 31 Serial Cable PC COM1 Port Spinstand Serial Port E5022-61628 1. Chapter Title Replaceable Parts Replaceable Parts Part No. 2. Chapter Title *1.
Replaceable Parts Replaceable Parts Cables and Connectors of E5023A Option 300 Figure 3-4 Cable Connection of E5023A Option 300 Table 3-6 Cable Connection (Option 300) No. Description Connection Part Number 34 BNC(m)-BNC(m) Cable, 2.3 m E5035B “Trig Out 2” E4402B “Gate Trig / Ext. Trig In” E5023-61615 35 BNC(m)-SMA(m) Cable, 2.
Table 3-6 No. Cable Connection (Option 300) Description Connection 1. Chapter Title Replaceable Parts Replaceable Parts Part Number BNC(m)-BNC(m) Cable, 2.3 m E5040A“Ref. Out” E4402B “10 MHz Ref in” E5023-61616 37 SMA(m)-SMA(m) Cable, 0.6 m E5036A “Thru Out 1” E4402B “Input” E5023-61617 38 GPIB Cable, 2 m GPIB Port on PCI Board E4402B “GPIB” 10833B 2. Chapter Title 36 3. Replaceable Parts 4. Chapter Title 5.
Replaceable Parts Replaceable Parts 84 Chapter 3
1. Chapter Title 2. Chapter Title 3. Chapter Title 4. Assembly Replacement and Post-Repair Procedures Assembly Replacement and Post-Repair Procedures 4 5.
Assembly Replacement and Post-Repair Procedures Introduction Introduction In this chapter, information required for replacement of the instruments in the Agilent E5023A test system is provided. Refer to this chapter when replacing or updating the instruments. Topics in this chapter are as follows.
Replacement of VXI Main Frame (Agilent E8401A) 1. Chapter Title Assembly Replacement and Post-Repair Procedures Replacement of VXI Main Frame (Agilent E8401A) This part describes how to replace the Agilent E8401A VXI main frame. Tools Required Pozidriv screwdriver • Torque Limiting Driver - 8 mm Opening 1.1 N⋅m (11 kgf⋅cm) 2. Chapter Title • Removal To uninstall the VXI main frame from the system, follow below steps. Step 1.
Assembly Replacement and Post-Repair Procedures Replacement of E8491B (IEEE 1394 PC Link to VXI Module) Replacement of E8491B (IEEE 1394 PC Link to VXI Module) This part describes how to replace the Agilent E8401A IEEE 1394 PC-Link-to-VXI module. Tools Required • Pozidriv screwdriver Removal To uninstall the E8491B module from the VXI mainframe, follow below steps. Step 1. Shut whole system down. Step 2. Remove all cables connected to the module. Step 3. Pull the module out from the VXI mainframe.
Replacement of E5035B/37C/37D/38B/40A/41A Modules 1. Chapter Title Assembly Replacement and Post-Repair Procedures Replacement of E5035B/37C/37D/38B/40A/41A Modules This part describes how to replace the E5035B/37C/37D/38B/40A/41A modules. Tools Required Pozidriv screwdriver • Torque Limiting Driver - 8 mm Opening 1.1 N⋅m (11 kgf⋅cm) 2. Chapter Title • Removal To uninstall the module from the VXI mainframe, follow below steps. Step 1. Shut whole system down. Step 2.
Assembly Replacement and Post-Repair Procedures Replacement of E5035B/37C/37D/38B/40A/41A Modules Step 4. Type the command shown below. This command will download the firmware to the module. It takes a few minutes. Module Command (number designates module’s logical address) E5035B update 5 ..\dsp [Enter] E5037C/D update 7 ..\dsp [Enter] E5038B update 8 ..\dsp [Enter] E5040A update 10 ..\dsp [Enter] E5041A update 11 ..\dsp [Enter] Step 5.
Replacement of E5036A Module 1. Chapter Title Assembly Replacement and Post-Repair Procedures Replacement of E5036A Module This part describes how to replace the E5036A module. Tools Required Pozidriv screwdriver • Torque Limiting Driver - 8 mm Opening 1.1 N⋅m (11kgf⋅cm) 2. Chapter Title • Removal To uninstall the module from the VXI mainframe, follow below steps. Step 1. Shut whole system down. Step 2. Remove all cables connected to the module. Step 3. Pull the module out from the VXI mainframe.
Assembly Replacement and Post-Repair Procedures Replacement of E5036A Module Figure 4-1 E5036A Diagram Operation Check Refer to the “Perform System Initialization” on page 28 to verify the system operation.
Replacement of E5039B Module 1. Chapter Title Assembly Replacement and Post-Repair Procedures Replacement of E5039B Module This part describes how to replace the E5039B module. Tools Required Pozidriv screwdriver • Torque Limiting Driver - 8 mm Opening 1.1 N⋅m (11 kgf⋅cm) • T10 TORX driver • T8 TORX driver • Socket Wrench 5/8 (Long Pattern)*1 • Open End or Box Wrench 5/16 • Hex Nut Driver 3/16 or 5mm *1 2. Chapter Title • Approximately 20mm depth is required. 3.
Assembly Replacement and Post-Repair Procedures Replacement of E5039B Module Replacement To install the module into the VXI mainframe, follow below steps. Step 1. Open the new module and install the channel IC board as following steps. 1. Install new channel IC board (No. 2) 2. Screw 3 trox screws (No. 9) 3. Replace 5 spacers (No. 15) on the SMA connectors of the channel IC board. 4. Replace the right side panel (No. 3). 5. Screw 2 jack screws. 6. Replace the front panel (No. 4). 7.
Figure 4-2 E5039B Diagram 2. Chapter Title 3. Chapter Title 4. Assembly Replacement and Post-Repair Procedures 5. Chapter Title 95 Chapter 4 1.
Assembly Replacement and Post-Repair Procedures Replacement of E5039B Module Post Replacement Procedures Following steps to update the module firmware are required after replacement of the modules. If you don’t follow these, “revision conflict error” will be occurred when you launch the system software. Step 1. Turn on and log in the PC. Step 2. Launch MS-DOS Prompt (Start - Programs - MS-DOS Prompt) from Start menu of the Windows 95 or 2000, and change the directory to c:\Program Files\Agilent\E5022\Util.
Replacement of E5039C Module 1. Chapter Title Assembly Replacement and Post-Repair Procedures Replacement of E5039C Module This part describes how to replace the E5039C module. Tools Required Pozidriv screwdriver • T10 TORX driver*1 • Wrench*1 2. Chapter Title • Removal To uninstall the module from the VXI mainframe, follow below steps. Step 1. Shut whole system down. Step 2. Remove all cables connected to the module. Step 4.
Assembly Replacement and Post-Repair Procedures Replacement of E5039C Module Step 1. Open the new module and install the channel IC board as following steps. 1. Remove 10 torx screws (No. 1 in Figure 4-3) fasten on the right side panel. 2. Remove the right side panel (No. 8). 3. Install the channel IC board (No. 9). 4. Tighten 4 torx screws (No. 2) fasten on the channel IC board. 5. Tighten 3 nuts (No. 3) and washers (No. 4) that secure the SMA connectors to the front panel. 6. Tighter 2 nuts (No.
Figure 4-3 E5039C Diagram 2. Chapter Title 3. Chapter Title 4. Assembly Replacement and Post-Repair Procedures 5. Chapter Title 99 Chapter 4 1.
Assembly Replacement and Post-Repair Procedures Replacement of E5039C Module Post Replacement Procedures Following steps to update the module firmware are required after replacement of the modules. If you don’t follow these, “revision conflict error” will be occurred when you launch the system software. Step 1. Turn on and log in the PC. Step 2. Launch MS-DOS Prompt (Start - Programs - MS-DOS Prompt) from Start menu of the Windows 95 or 2000, and change the directory to c:\Program Files\Agilent\E5022\Util.
Replacement of Oscilloscope (Agilent 54845A/B) or Spectrum Analyzer (Agilent E4402B) 1. Chapter Title Assembly Replacement and Post-Repair Procedures Replacement of Oscilloscope (Agilent 54845A/B) or Spectrum Analyzer (Agilent E4402B) This part describes how to replace the optional oscilloscope/spectrum analyzer. Removal 2. Chapter Title To uninstall the oscilloscope/spectrum analyzer from the system, follow below steps. Step 1. Shut whole system down. Step 2.
Assembly Replacement and Post-Repair Procedures Replacement of Onstage Buffer Unit and Connection Board Replacement of Onstage Buffer Unit and Connection Board Overview Head amplifier is a module that enlarges signals from the hard disk head. Head amplifier differs depending on what IC is used in the hard disk that the head is going to be installed. This section describes a method to replace a head amplifier. Tools • T10 TORX driver Removal of Head Amplifier Step 1. Shut whole system down. Step 2.
contact to the electrode pattern on the head amplifier. If contact is bad, make a fine adjustment on the positioning of the head amplifier after attachment. 1. Chapter Title Assembly Replacement and Post-Repair Procedures Replacement of Onstage Buffer Unit and Connection Board Step 3. Fasten screws and fix the head amplifier to the HLM tightly. Inspecting the Pads Condition Four sets of this are furnished with the system as spares. When you need them more, order PN E5029-65001.
Assembly Replacement and Post-Repair Procedures PC (System Controller) Replacement PC (System Controller) Replacement Tools Required • Pozidriv screwdriver Removal Step 1. Remove all cables (GPIB cable, IEEE-1394 cable, and serial cable) from the PC. Step 2. Remove GPIB card from the PC. Step 3. Remove IEEE-1394 card from the PC. Replacement Step 1. Install the Agilent E5023A software including Agilent I/O Library into the PC.
GPIB Card and/or IEEE-1394 Card Replacement 1. Chapter Title Assembly Replacement and Post-Repair Procedures GPIB Card and/or IEEE-1394 Card Replacement Tools Required • Pozidriv screwdriver 2. Chapter Title Removal Step 1. Remove all cables (GPIB cable, IEEE-1394 cable, and serial cable) from the PC. Step 2. Remove GPIB card and/or IEEE-1394 card from the PC. Replacement Step 1. Install a GPIB card and/or an IEEE-1394 card into slot(s) of the PC.
Assembly Replacement and Post-Repair Procedures Cable Connection Cable Connection Cable Connection of E5023A Option 415 (1.5 Gbps) (without opt. 300) Figure 4-5 shows the overall connection diagram including optional oscilloscope and the E5039B/C bit error test module. The E5041A dual counter module is not included in this figure.
Figure 4-5 Overall Connection Diagram of E5023A Option 415 (1.5 Gbps) (without opt. 300) 1. Chapter Title Assembly Replacement and Post-Repair Procedures Cable Connection 2. Chapter Title 3. Chapter Title 4. Assembly Replacement and Post-Repair Procedures 5.
Assembly Replacement and Post-Repair Procedures Cable Connection Figure 4-6 shows the cable connection of the head amplifier and the HLM. Figure 4-6 Cable Connection for E5023A-415 (1.5 Gbps) (Connection Board, Onstage Buffer Unit, E5043A) NOTE The locations of connectors on the head amplifier may differ from the figure above depending on the amplifier that you use. Check the connector names printed on the head amplifier when connecting the cables.
Table 4-1 No. Cable Connection (Head Amplifier, HLM) (E5023A-415, 1.5 Gbps) Description Connection 1.
Assembly Replacement and Post-Repair Procedures Cable Connection Cable Connection of E5023A Option 426 (2.6 Gbps)(without opt. 300) Figure 4-7 shows the overall connection diagram including optional oscilloscope and the E5039B/C bit error test module. The E5041A dual counter module is not included in this figure.
Figure 4-7 Cable Connection of E5023A Option 426 (2.6 Gbps) (without opt. 300) 1. Chapter Title Assembly Replacement and Post-Repair Procedures Cable Connection 2. Chapter Title 3. Chapter Title 4. Assembly Replacement and Post-Repair Procedures 5.
Assembly Replacement and Post-Repair Procedures Cable Connection Figure 4-8 shows the cable connection of the head amplifier and the HLM. Figure 4-8 Cable Connection for E5023A-426 (2.6 Gbps) (Connection Board, Onstage Buffer Unit, E5043C) NOTE The locations of connectors on the head amplifier may differ from the figure above depending on the amplifier that you use. Check the connector names printed on the head amplifier when connecting the cables. Table 4-2 Cable Connection of E5023A Opt. 426 (2.
Table 4-2 Cable Connection of E5023A Opt. 426 (2.
Assembly Replacement and Post-Repair Procedures Cable Connection Connecting the VXI Modules Figure 4-9 E8491B/E5035B Cable Connection *1 Already connected inside of the spinstand to the piezo control unit and the piezo power supply. Table 4-3 No.
E5037C Cable Connection (for E5023A-415, 1.5 Gbps) Table 4-4 E5037C Cable Connection (for E5023A-415, 1.5 Gbps) 2. Chapter Title Figure 4-10 1. Chapter Title Assembly Replacement and Post-Repair Procedures Cable Connection 3. Chapter Title No.
Assembly Replacement and Post-Repair Procedures Cable Connection Figure 4-11 E5037D Cable Connection (for E5023A-426, 2.6 Gbps) Table 4-5 E5037D Cable Connection (for E5023A-426, 2.6 Gbps) No. 11 Description Connection Part No.
Cable Connection (E5036A/E5038B) Table 4-6 E5036A/E5038B Cable Connection 2. Chapter Title Figure 4-12 1. Chapter Title Assembly Replacement and Post-Repair Procedures Cable Connection Description 3. Chapter Title No.
Assembly Replacement and Post-Repair Procedures Cable Connection Connecting the E5039B Bit Error Test Module (Optional) NOTE This procedure is needed only if the system is configured with the Agilent E5039B Bit Error Test Module. Figure 4-13 Cable Connection of the Agilent E5039B (Option) Table 4-7 Cable Connection (E5039B) No.
Connecting the E5039C Bit Error Test Module (Optional) This procedure is needed only if the system is configured with the Agilent E5039C Bit Error Test Module. Figure 4-14 Cable Connection of the Agilent E5039C (Option) Table 4-8 Cable Connection (E5039C) 2. Chapter Title NOTE 1.
Assembly Replacement and Post-Repair Procedures Cable Connection Connecting the E5040A Spectrum Analyzer Module Figure 4-15 Cable Connection (Spectrum Analyzer Module) Table 4-9 Cable Connection (E5040A) No.
Connecting the E5041A Dual Counter Module (Optional) This procedure is needed only if the system is configured with the Agilent E5041A Dual Counter Module. Figure 4-16 Cable Connection of the Agilent E5041A (Option) Table 4-10 Cable Connection (E5041A) 2. Chapter Title NOTE 1. Chapter Title Assembly Replacement and Post-Repair Procedures Cable Connection 3.
Assembly Replacement and Post-Repair Procedures Cable Connection Connecting Spinstand and PC Figure 4-17 Cable Connection (Spinstand) 122 Chapter 4
Cable Connection (PC) (without option 300) Table 4-11 Cable Connection (Serial Cable) 2. Chapter Title Figure 4-18 1. Chapter Title Assembly Replacement and Post-Repair Procedures Cable Connection 3. Chapter Title No.
Assembly Replacement and Post-Repair Procedures Cable Connection Terminating unused connectors Figure 4-19 Terminator Connection Table 4-12 Terminator Connection No.
Connecting the Oscilloscope (Option) Connect the oscilloscope as follows. 1. Chapter Title Assembly Replacement and Post-Repair Procedures Cable Connection Connecting Option 103 Connect the cables for option 103 (8 GSa/s Oscilloscope) as shown in Figure 4-20 and Table 4-13. 2. Chapter Title 3. Chapter Title 4. Assembly Replacement and Post-Repair Procedures 5.
Assembly Replacement and Post-Repair Procedures Cable Connection Figure 4-20 Cable Connection (Option 103) Table 4-13 Cable Connection (Option 103) No.
Table 4-13 No. Cable Connection (Option 103) Description Connection Part Number 23 SMA(f)-BNC(m) Adapter - - 1250-1700 27 GPIB Cable, 2m GPIB Port on PCI Board 54845A/B GPIB Port 10833B 28 GPIB Adapter Extender*1 - - 10834A Store the terminators you removed from the Agilent E5035B to mount the oscilloscope in a safe place for future use. 2. Chapter Title *1.The GPIB Extender is used if the GPIB connector is difficult to connect to the PC when the PC’s rear panel is sunk. NOTE 1.
Assembly Replacement and Post-Repair Procedures Cable Connection Connecting the 3 GHz Spectrum Analyzer (Option) Connect the 3 GHz Spectrum Analyzer as follows. Connecting Option 300 Connect the cables for option 300 (3 GHz Spectrum Measurement Capability) as shown in Figure 4-21 and Table 4-14.
Cable Connection (Option 300) Table 4-14 Cable Connection (Option 300) 2. Chapter Title Figure 4-21 1. Chapter Title Assembly Replacement and Post-Repair Procedures Cable Connection 3. Chapter Title Description Connection Part Number BNC(m)-BNC(m) Cable, 2.3 m E5035B “Trig Out 2” E4402B “Gate Trig / Ext. Trig In” E5023-61615 35 BNC(m)-SMA(m) Cable, 2.3 m E5040A “Signal In” E4402B “Aux IF Out” E5023-61618 36 BNC(m)-BNC(m) Cable, 2.3 m E5040A“Ref.
Assembly Replacement and Post-Repair Procedures Cable Connection Table 4-14 No. 38 Cable Connection (Option 300) Description GPIB Cable, 2 m Connection GPIB Port on PCI Board E4402B “GPIB” Part Number 10833B Connecting the GPIB Step 1. Connect a GP-IB extender adapter to the GPIB connector of the PC. Step 2. Connect the PC and the E4402B with a GPIB cable.
How to Check the Operation After Replacement 1. Chapter Title Assembly Replacement and Post-Repair Procedures How to Check the Operation After Replacement After replacement of the instrument, it is necessary to perform the software updating, re-configuration and re-adjustment as shown in Table 3-11. And, perform the System Check function or user application program again to verify the E5023A operation. Table 4-15 Requirements after replacement of instruments (---: not required) s/w update re-config.
Assembly Replacement and Post-Repair Procedures How to Check the Operation After Replacement *6.If exchanged head cassette is not same as existed one, it is necessary to re-enter the cassette parameter. *7.After replacing an assembly of spinstand system, you may additionally need to perform re-adjustment. See spinstand service manual for details. It is able to confirm the E5023A system configuration, such as instrument serial number, software revisions and driver revisions.
1. Chapter Title 2. Chapter Title 3. Chapter Title 4. Chapter Title 5 Software Recovery 133 5. Software Recovery This chapter provides the information on how to install the required software for the PC of the E5023A. When the PC breaks down and is replaced with a new one, you will have to install the E5023A software into the PC.
Software Recovery Installing Software Installing Software The E5023A system requires some software to be installed into the PC. The following sections provide the important information when you install the following software into the PC. NOTE • Agilent I/O Library • Agilent VEE • Acrobat Reader • Agilent E5023A System Software. The Agilent E5023A system supports Windows 95 or Windows 2000 operating system. Windows 98/ME/NT are not supported.
Installing Agilent I/O Library 1. Chapter Title Software Recovery Installing Agilent I/O Library Agilent I/O library is a general I/O library that controls the instrument. This includes the Agilent VISA / SICL library. The E5023A system software requires this software to control VXI module, instruments such as an oscilloscope, and spinstand through IEEE-1394, GPIB and RS-232C. Follow the instruction on the screen to install the I/O Library, basically.
Software Recovery Installing Agilent I/O Library 4. Read the readme information and click Next >. 5. If required, change the SICL installation holder using Browse.... Then click Next >. 6. If required, change the VISA installation holder using Browse.... Then click Next >.
7. Select 1. SICL and Agilent VISA Installation and click Next >. 1. Chapter Title Software Recovery Installing Agilent I/O Library 2. Chapter Title NOTE If you are asked if you want to install Agilent E8491 IEEE 1394 to VXI support, you must click Yes. 8. Click Next > to start copying the program files. 3. Chapter Title 9. Select Auto-configure all interfaces and click Next >. 4. Chapter Title 5.
Software Recovery Installing Agilent I/O Library 10. Click Finish to finish the installation. Configuring I/O Library 1. Connect the E8491B with the IEEE-1394 card in the PC slot. 2. Turn on the PC and the VXI mainframe. 3. Execute the I/O Config by clicking the right mouse button on the IO logo in the windows task bar (right bottom in the PC screen) and selecting Run IO Config. 4. Click Auto Add. 5. Confirm if the COM1, hpib7 and VXI is displayed as the SICL Name in the Configured Interfaces box.
6. Select COM1 ASRL1 in the Configured Interfaces, then click Edit. 1. Chapter Title Software Recovery Installing Agilent I/O Library 2. Chapter Title 7. Set the parameters as shown below. • Baud Rate: 19200 • Flow Control: None 3. Chapter Title 8. Click OK. 4. Chapter Title 5.
Software Recovery Installing Agilent I/O Library 9. Click OK. 10. Reboot the PC. Checking Connection 1. Turn on the PC and the VXI mainframe. 2. Execute the VISA assistant by clicking the right mouse button on the IO logo in the task bard (right bottom in the PC screen) and select Run VISA Assistant. 3. Confirm if the ASRL1, VXI0, GPIB0 are displayed on the left side of the screen. NOTE The following property might not be displayed when module is not installed in the VXI.
• VXI0:6:INSTR (E5036A) • VXI0:7:INSTR (E5037C/D) • VXI0:8:INSTR (E5038B) • VXI0:9:INSTR (E5039B/C) • VXI0:10:INSTR (E5040A) • VXI0:11:INSTR (E5041A) a. Select GPIB0::7::INSTR in the left side of the screen. b. Select Formatted I/O tab. c. Check IEEE 488.2. 2. Chapter Title 4. Confirm if the GPIB control properly works if the 54845A Oscilloscope is connected. 1. Chapter Title Software Recovery Installing Agilent I/O Library d. Click *RST. e.
Software Recovery Installing Agilent VEE Installing Agilent VEE Agilent E5023A VEE (Visual Engineering Environment) is a graphical programming language designed to solve engineering problems and create test systems. The Agilent VEE has two basic parts, namely; the development environment that allows you to use all features of the VEE and the run-time environment that allows you to run the VEE program created from the development environment.
4. Click Yes if you accept the terms of the license agreement. 2. Chapter Title 5. Enter your name, company name, and serial number of the software and click Next >. 1. Chapter Title Software Recovery Installing Agilent VEE 3. Chapter Title 6. Click Next >. 4. Chapter Title 5.
Software Recovery Installing Agilent VEE 7. Click Next >. 8. Select Typical and click Next >. 9. If you want to run the Agilent VEE now, select Yes, I want to run Agilent VEE. Click Finish to complete the installation. VEE RunTime Environment Installation Procedure 1. Insert the Agilent VEE program CD into the CD-ROM drive of the PC. Agilent VEE installation program will start automatically. 2.
Environment. 1. Chapter Title Software Recovery Installing Agilent VEE 2. Chapter Title 3. Click Next >. 3. Chapter Title 4. Click Yes if you accept the terms of the license agreement. 4. Chapter Title 5.
Software Recovery Installing Agilent VEE 5. Click Next >. 6. Select Typical and click Next >. 7. Click Finish to complete the installation.
Installing Acrobat Reader 1. Chapter Title Software Recovery Installing Acrobat Reader Acrobat Reader allows you to read the operation/programming manuals in pdf format. It is provided in the E5023A system software CD-ROM or the website of Adobe Company (http:\\www.adobe.com\). Follow the instruction of the installer program. Installation Procedure 2. Chapter Title 1. Click Next >. 3. Chapter Title 2. Click Next >. 4. Chapter Title 3. Click OK. 5.
Software Recovery Installing Acrobat Reader 4. Click Accept if you accept the terms of the license agreement.
Installing E5023A System Software 1. Chapter Title Software Recovery Installing E5023A System Software The E5023A system software allows you to control the E5023A. Executing the system software installs the E5023A function library, VEE measurement software, manuals in PDF format. It is provided in the E5022/E5023 system software CD-ROM or the website of Agilent E5023A customer support (http://dst.tm.agilent.com/). 1.
Software Recovery Installing E5023A System Software 6. If Shared File Detected dialog box appears, select Don’t display this message again and click No. 7. If ReadOnly File Detected dialog box appears, select Don’t display this message again and click Yes. 8. Click Finish to end the system software removal. 9. Select and double-click the system software: inst_e5022.exe again. 10. Click Next > to continue the installation.
11. Read the information and click Next >. 2. Chapter Title 12. If needed, change the system installation holder using Browse.... Then click Next >. 1. Chapter Title Software Recovery Installing E5023A System Software 3. Chapter Title 13. Enter the system’s serial number and click Next >. 4. Chapter Title 14. If firmware update is needed, select the desired firmware of VXI modules. You can use Select All or Clear All to select all items or clear all selections at a time. 5.
Software Recovery Installing E5023A System Software NOTE Do not select a firmware check box if its hardware is not installed into the system. 15. Click Next > to start firmware update. 16. If the head amplifier/cassette configuration software should be installed, select Install Headamp/Cassette Configuration and click Next >.
17. Click Next > to start installing the Head Amplifier/Cassette Configuration Software. 2. Chapter Title 18. If ReadOnly File Detected dialog box appears, select Don’t display this message again and click Yes. 1. Chapter Title Software Recovery Installing E5023A System Software 3. Chapter Title 19. Click Finish. 4. Chapter Title 20. Select Yes, I want to restart my computer now, remove the CD-ROM from the PC, and click Finish to restart the computer. 5.
Software Recovery PC Power Management PC Power Management If your E5023A has Windows 95 operating system, you must turn off the power management mode to avoid performance (measurement speed) deterioration. For Windows 2000 operating system shipped with the latest version of the E5023A, you don’t need to care the power management mode because Windows 2000 does not have such mode. Use the ‘Power Icon’ in the Control Panel to change this setting.
A. Manual Changes 2. Chapter Title 3. Chapter Title 4. Chapter Title A Manual Changes 155 5. Chapter Title This appendix contains the information required to adapt this manual to versions or configurations of the E5023A manufactured earlier than the current printing date of this manual. The information in this manual applies directly to E5023A units having the serial number printed on the title page of this manual.
Manual Changes Manual Changes Manual Changes To adapt this manual to your E5023A, refer to Table A-1 and Table A-2. Table A-1 Manual Changes by Serial Number Serial Prefix or Number Table A-2 Make Manual Changes Manual Changes by Firmware Version Version Make Manual Changes Agilent Technologies uses a two-part, ten-character serial number that is stamped on the serial number plate (Figure A-1). The first five characters are the serial prefix and the last five digits are the suffix.
1. Chapter Title B. Measurement Paths 3. Chapter Title 4. Chapter Title B Measurement Paths 157 5. Chapter Title In this Appendix, major measurement parameters versus signal paths are described. The matrix is useful to isolate faulty part or module if measurement parameter value has some problem.
Measurement Paths Measurement Paths by Measurement Parameters Measurement Paths by Measurement Parameters Table B-1 E5035B Measurement Paths by Typical Measurement Parameters Module E5035B Spinstand Interface Module Function Trigger Output Port Head Amp Head Amp Trig Out 1 VXI bus (to Head Amp (to Osc) 37/38/39/40) Ctrl Unit I/F Ctrl Unit I/F Ctrl Unit I/F (from 43) (to 43) (to 43) Write Current Control Sense Current Control DCR Write Gate Measurement Control Head Amp Control Piezo Stage Contro
Table B-3 E5038B/E5040A Measurement Paths by Typical Measurement Parameters Module E5038B Parametric Measurement Module Function Trigger Acceptance Port VXI bus (from Signal In (from Signal In (from Signal In (from VXI bus (from 35) 36) 36) 36) 35) TAA P PW P Baseline P TAA Measurement E5040A Spectrum Measurement Module PW Measurement Baseline Measurement Trigger Acceptance External Reference Signal Acceptance Spectrum Measurement Ext Ref In (from 37) Signal In (from 36) M M NLTS 5th
Measurement Paths Measurement Paths by Measurement Parameters Table B-5 E4402B Measurement Paths by Typical Measurement Parameters Module E4402B Spectrum Analyzer (Optional) Function IF Signal Trigger External Output Acceptance Reference Signal Acceptance Port EXT TRIG 10 MHz REF IN IN (from (from 40) 35) AUX IF INPUT (from OUT (to 40) 36) NLTS 5th P P P P Narrow Band TAA P P P P P P P P P P P M Spectrum Measurement TAA PW Baseline Triple Track Profile (TAA) Over Write DCR (Rea
1. Chapter Title 2. Chapter Title C. Software Tools 4. Chapter Title C Software Tools 5. Chapter Title This appendix describes how to use software tools installed in the Agilent E5023A.
Software Tools Introduction Introduction This appendix describes how to use software tools installed in the Agilent E5023A. These tools are used to troubleshoot and configure the setting after module replacement. The items shown below are explained in the following sections.
Spinstand Tool Usage 1. Chapter Title Software Tools Spinstand Tool Usage Purpose of Tools The spinstand tools allows you to control the each stage, spindle and HLM one by one to help adjustment, repair and isolation. This tool has following function. Spinstand Data Backup/Restore Backup or restore the spinstand data. This data includes important and unique values for each spinstand. So, you should back up the data before every adjustment/repair for the spinstand. • Spinstand Diagnosis 2.
Software Tools Spinstand Tool Usage Procedure to use the Spinstand Tool The procedure for the spinstand tool is as follows. Step 1. Connect the spinstand and the PC with serial cable. Step 2. Turn on the spinstand and the PC. Step 3. Select Start - Programs - Agilent Hard Disk ReadWrite Test System - Spinstand Tool to execute the spinstand tool NOTE It is strongly recommended to reboot the PC before launching spinstand tool to free serial port from other applications.
Figure C-2 Spinstand Data Backup/Restore menu 1. Chapter Title Software Tools Spinstand Tool Usage 2. Chapter Title Backup all spinstand data to the PC. To back up the data, click this button and select the destination of the data. Restore Restore the backuped data to the spinstand. To restore the data, click this button and select the file you want to restore. While backing up and restoring the data, the Spinstand Data Backup/Restore menu may be corrupted.
Software Tools Spinstand Tool Usage Spinstand Diagnosis menu To activate Spinstand Diagnosis menu, click on Diagnostic tab of the spinstand tool. This menu controls each part of the spinstand to diagnosis the spinstand. Status window shows each part status and data. Figure C-3 Spinstand Diagnosis menu Reset Reset the spinstand. Axis Select Select an Axis from Spindle Stage, HLM Stage, Spindle Motor or All Motors. Stage Lock Lockdown the selected stage. Stage Unlock Unlock the selected stage.
Figure C-4 Spinstand Diagnosis menu Status window 1. Chapter Title Software Tools Spinstand Tool Usage 2. Chapter Title C. Software Tools 4. Chapter Title 5.
Software Tools Usage of I/O Config Usage of I/O Config Purpose of I/O Config I/O Config is to detect the interface devices and configure them. The I/O config is set at proper setting at factory shipment. However, somehow it might be change. In case of communication trouble, confirming I/O configure is required. Procedure of I/O Config The procedure for I/O Config is as follows. Step 1.
VISA Assistant Usage 1. Chapter Title Software Tools VISA Assistant Usage Purpose of VISA Assistant VISA Assistant is designed to help you communicate with instruments. VISA Assistant can automatically detect and assign Instrument Drivers to Instruments. In case of communication trouble, it allows you to confirm if the communication is correct. Step 1.
Software Tools VISA Assistant Usage Figure C-7 VISA Assistant Menu Step 4. Confirm as same manner for "VXI::0::INSTR","VXI::5::INSTR", "VXI::6::INSTR", "VXI::7::INSTR", "VXI::8::INSTR", "GPIB0::xx::INSTR", "ASRL1::INSTR". If the connection has problem, they are different between the selection of right side screen and the displayed in the middle of the left side screen. See the Figure C-8 on page 170. Figure C-8 Connection Problem Example Step 5.
quit the resource manager. 3. Select Start - Programs - StartUP - Agilent I_O Libs Control from Start menu of Windows 95 to execute the resource manager again. 1. Chapter Title Software Tools VISA Assistant Usage 4. Click the right mouse button on the HP logo in the task bar and select Refresh Resource Manager to refresh the status of the instruments. Step 6. Confirm if the connection makes correct according to the step 1 to 3. 2. Chapter Title Step 7.
Software Tools VISA Assistant Usage 172 Appendix C
1. Chapter Title 2. Chapter Title 3. Chapter Title D. Error Messages Error Messages D 5. Chapter Title This appendix lists the error messages.
Error Messages Error Messages on PC Monitor Display Error Messages on PC Monitor Display Table shows the error messages that could appear on the PC monitor display when you use a demo program or other programs created by the user. Error Messages on PC Monitor Display Error Code Description hpe5022_ERROR_AUTO_CONFIG_FAILED The auto configuration has failed. Check if the parameter and head is correct. hpe5022_ERROR_AUTO_RANGE_FAIL The auto range has failed. Check if the parameter and head is correct.
Error Messages on PC Monitor Display hpe5022_ERROR_FUNCTION_NOT_LOADED Unable to load the DLL file. Perform the E5023A installation software or check if the file is installed in the PC. The location of DLL file is described in Chapter 2 of the programming manual. hpe5022_ERROR_HARD Detected a hardware error in the system. hpe5022_ERROR_HARD_HAMP Detected a hardware in the head amplifier board. hpe5022_ERROR_HARD_MISS The media or HGA cassette is missing or the selected hardware is not installed.
Error Messages Error Messages on PC Monitor Display Error Messages on PC Monitor Display Error Code Description hpe5022_ERROR_INV_DATA_SIZE In the user defined data pattern, the length of specified data pattern is shorter than the specified user data bit length. hpe5022_ERROR_INV_DATA_TYPE The specified data pattern for the user define data pattern includes an invalid character.
Error Messages on PC Monitor Display hpe5022_ERROR_INV_SEQ_HNDL Detected the invalid sequence handle parameter in UDS. Check your program. hpe5022_ERROR_INV_SETUP Detected an error in operation. There must be some illegal set-up in the parameters. Check your program. hpe5022_ERROR_INV_TRIG_PORT Invalid trigger port is selected as the parameter ‘trigPort’. hpe5022_ERROR_MEM_ALLOC Can't allocate enough memory. Reboot your PC.
Error Messages Error Messages on PC Monitor Display Error Messages on PC Monitor Display Error Code Description hpe5022_ERROR_NSUP_CHAN_IC The DLL for the installed channel IC is not installed. Install the E5039B/C option software. hpe5022_ERROR_NSUP_CONF Detected software configuration errors. Check your program. hpe5022_ERROR_NSUP_FUNC Detected a mismatch between measurement functions, firmware and hardware. Check the revisions of those.
Error Messages on PC Monitor Display The head does not support the reverse polarity on a stability measurement. Set the polarity at normal. hpe5022_ERROR_OPEN Unable to open some components in initialization. (instruments and the spinstand). hpe5022_ERROR_OPT_FAIL The optimization is failed. hpe5022_ERROR_OVERFLOW An overflow is detected in the parametric module. Perform the auto configuration (input range) or Check if the setting parameter is correct.
Error Messages Error Messages on PC Monitor Display Error Messages on PC Monitor Display Error Code Description hpe5022_ERROR_RESET Unable to reset on some components (instruments and the spinstand). hpe5022_ERROR_SERCH_TRACK_FAILED Detected an error and failed the operation in searching the track. hpe5022_ERROR_SEG_COMPLETE Segment is already completed in UDS. Check your program. hpe5022_ERROR_SEG_IN_EDIT Segment is not completed in UDS. Check your program.
Error Messages on PC Monitor Display hpe5022_ERROR_TOO_MANY_RETRY Spinstand stage is not positioned within a certain range after retry. Call Agilent Service office. hpe5022_ERROR_TRACK_OFFSET_COMP_FAILE D Track offset compensation is failed. Check if the all measurement conditions are correct. hpe5022_ERROR_UNAVAILABLE_PINNO Unavailable pin number is assigned for the S/S TTL I/O. Check the available pin number. hpe5022_ERROR_UNDERFLOW An underflow is detected in the parametric module.
Error Messages Error Messages at Auto Disk Alignment Error Messages at Auto Disk Alignment Table D-1 shows the error messages that could appear when Auto Disk Alignment is performed. Table D-1 Error Message at Auto Disk Alignment Error Message Description Can't converge the track position The start position could be wrong or the X-Y positioning might be inaccurate. Retry auto disk alignment. When it fail again, contact Agilent Technologies.
Index Numerics 4395A, 31 54845A, 101 82350 GPIB, 168 A acrobat reader, 147 Agilent I/O Library, 20 Agilent part number, 69 Agilent VEE, 21 Agilent VISA Assistant, 31 Agilent VISA library, 20 all measurement items, 65 ASRL1, 31, 168 ASRL1:: INSTR:BFFF0011, 32, 33 ASRL1::INSTR, 169, 170 ASRL1::INSTR:0x80001106, 33 ASRL1::INSTR:0x8000110a, 33 ASRL1::INSTR:80001108, 33 ASRL1::INSTR:BFFF0015, 32 ASRL1::INSTR:BFFF003E, 32 ASRL1::INSTR:xxxxxxxxx, 33 Auto Configuration, 63 Auto Configuration Error, 63 B BER Measur
Index hpe5022_ERROR_INV_POSITION, 176 hpe5022_ERROR_INV_SEQUENCE, 176 hpe5022_ERROR_INV_TRIG_PORT, 176, 177 hpe5022_ERROR_MEM_ALLOC, 177 hpe5022_ERROR_MEMORY, 29 hpe5022_ERROR_MICRO_TRACK_FAILED, 177 hpe5022_ERROR_MOTOR_FAULT, 33, 177 hpe5022_ERROR_NDEF_HGA, 177 hpe5022_ERROR_NO_OPTION, 31, 177 hpe5022_ERROR_NSUP_CHAN_IC, 178 hpe5022_ERROR_NSUP_HAMP, 178 hpe5022_ERROR_NSUP_HGA_CASSETTE, 178 hpe5022_ERROR_NSUP_OPER, 178 hpe5022_ERROR_NSUP_SENS_CURR_POL, 179 hpe5022_ERROR_OPT_FAIL, 179 hpe5022_ERROR_OVERFLOW
Index T Timeout occured, 64 Track Profile, 64 U Unclamp, 163 Unlock, 163 V vee, 142 VI_ERROR_IO, 32 VI_ERROR_PRES_NFOUND, 29 VI_ERROR_RSRS_NFOUND, 29, 31, 32, 33 VISA Assistant, 169 VISA name, 31, 168 Visual Basic, 21 VXI Main Frame, 26, 27, 87, 131 VXI::0::INSTR, 169, 170 VXI::5::INSTR, 169, 170 VXI::6::INSTR, 169, 170 VXI::7::INSTR, 169, 170 VXI::8::INSTR, 169, 170 VXI0, 168 VXI0::0::INSTR, 31 VXI0::5::INSTR, 31 VXI0::5::INSTR:80001105, 29 VXI0::5::INSTR:80001106, 29 VXI0::6::INSTR, 31 VXI0::7::INSTR, 31