All other slots blocked. Airflow decreases as additional slots are
opened.
Performance shown for Worst Slot (slot 2). Airflow is greater in all
other slots.
Fans on Full Speed. Minimum airflow is approximately 50% with
fans on Variable Speed.
Air Filter Kit not installed. Airflow is reduced approximately 10%
with clean air filters installed.
Measurements taken at 1,500m altitude.
Cooling Mode (High or
Variable)
Switchable on the front panel. Controls both impeller and fan.
High Fan Speed Mode: Full Airflow all the time.
Variable Fan Speed Mode:
Fan Speed increments through 5 discrete speeds as a function of power
supply temperature and reference temperature. Reference
temperature (pins 11,12 of Diagnostic Connector) is measured on
the backplane and is a function of load and ambient temperature.
At full load: Low Speed up to approximately 30°C ambient.
High Speed above approximately 40°C ambient.
At no load: Low Speed up to approximately 40°C ambient.
High Speed above approximately 50°C ambient.
Assuming default limits:
At full load: Low Speed up to approximately 30°C
High Speed above approximately 40°C
At no load: Low Speed up to approximately 40°C
High Speed above approximately 50°C
18.0
12.0
6.0
24.0
5.0 10.0 15.0 20.0 25.0
Airflow Through Module (liters/sec)
0.0
Worst Slot
Pressure Drop
Across Module
(mm H
2
0)