Service manual

54 HP E8401A, E8403A Product Specifications Appendix A
Output Power Specifications
Total Available and
Usable Power
Peak and Dynamic
Current
Product Temperature Available Power
a
90-264VAC
Usable Power
b
110-264VAC
Usable Power
b
90-110VAC
E8401A 0-55°C 686 W 500 W 500 W
E8403A 0-55°C 1,902 W 1,000 W 950W
a.Sum of voltages times currents. Not always usable due to thermal protection
shutdown.
b.Total output before thermal protection shutdown or safety limitation.
E8401A E8403A
Voltage
Peak Current,
I
MP
a
,
b
Dynamic Current,
I
MD
a
,
c
Peak Current,
I
MP
a
,
b
Dynamic Current,
I
MD
a
,
c
+5V 50A 5A 90A 9A
+12V 6A 1A 15A 2.5A
-12V 4A 1A 15A 2.5A
+24V 4A 2A 15A 5A
-24V 4A 2A 15A 5A
-5.2V 20A 2A 60A 8A
-2V 10A 1A 30A 5A
a.
Specifications apply at the backplane, 0-55
°
C.
b.
IMP = Rated mainframe peak DC output current as defined by the VXIbus Specification.
c.
IMD = Rated mainframe peak-to-peak dynamic current as defined in the VXIbus Specification by