Specifications

4.1.2 Configure the user’s PC for DCOM access to
the PNA application
From your PC:
1. Using your PC's Windows Explorer, find the analyzer’s
mapped local hard drive and click the Program
Files/Agilent/Network Analyzer/Automation
folder.
2. Double-click Automation.vbs and follow the
instructions. After this procedure is completed, a
Network Analyzer Setup icon is setup on your PC
desktop.
NOTE: Your PC’s operating system may require a
Visual Basic Scripting engine to run the .vbs file. To
download a free copy of a Visual Basic Scripting
engine, visit Microsoft’s
®
web site.
Problems?
When the network analyzer application is installed on
the instrument, the application and all the objects it
serves are registered on that machine. The instrument
should be ready to perform as a remote server.
If the analyzer is not working properly, try re-registering
the application and associated libraries through the fol-
lowing steps:
NOTE: The following must be done in the listed order
From the PNA network analyzer:
Find E835x.exe in C:\Program Files\Agilent\Network
Analyzer
1. Select and right-click 835x.exe
2. Select Register COM Server to register the
application.
3. Select and right-click 835xps.DLL
4. Select Register COM Server
DCOM settings
5. Click Start
6. Select RUN and type dcomcnfg
7. From the application list, find and select 835X
8. Click Properties
9. Select the Identity tab.
10. Select INTERACTIVE USER
11. Select the Security tab
12. Set permissions and access to CUSTOM
13. Edit the lists to specify the remote user
9