Burst shape files must have two or more points for both the rise
and the fall shape specification.
702 Load/Store Error; Invalid modulation file.
The user modulation file is invalid. The file is either corrupt or
is not a modulation file.
702 Load/Store Error; Unrecognized FIR file version.
The file is either corrupt or is not a FIR file.
702 Load/Store Error; Unrecognized modulation file version.
The user modulation file has an unrecognized version. The file is
either corrupt or is not a modulation file.
702 Load/Store Error; Unrecognized shape file version.
The Burst Shape file is corrupted, or is not in the proper format.
703 Feature not supported; File '{name}' Header Parameter Sample Rate is
specified as %gsps, but it must be in the range of 1sps to 100Msps.
This file will not work correctly on this platform.
The first {name} is the file that generated the error. The %g is the
value of the Sample Rate as specified in the file header.
703 Feature not supported; RS232 flow control is not supported.
Command is not supported for RS232 communication.
703 Feature not supported; The selected waveform was generated using an
option not present on this instrument. '{name}' Use Header Utilities
and View License Details for more information.
This error will be generated when a selected file requires an
option that is currently not licensed on this instrument. {name} is the
filename.
703 Feature not supported; Software update required: Please download the
latest application from http://www.agilent.com/find/signalstudio.
Missing parameter. When the state is on, Uplink Pilot requires all
following fields: power in db, SYNC code, and time offset.
704 Span adjusted due to selected sweep time; Choose a wider span or
shorter sweep time to avoid this conflict.
*** No documentation ***
768 Header value bumped; Header File Field {name} Bumped from {name} to
{name}.
This error is reported when trying to apply the header parameters
to the instrument state. If one of the parameters is not possible
on for this instrument, it will be "bumped" to the next appropriate
value. The first {name} will describe which header field contained the
bumped value (the second {name}). The third {name} is the new value
used.
800 Digital module configuration error; Could not configure FPGA(s) on the