User manual

lpErrorDoc.txt
First {name} is file name, second is file revision and third is
supported revision.
629 File format invalid; Sequence contains an invalid single waveform
sequence.
This is reported when an invalid single waveform sequence is found
in a sequence file.
629 File format invalid; Sequence contains an invalid waveform.
This is reported when an invalid waveform is found in a sequence
file.
629 File format invalid; Sequence references a waveform that is not
present. {name}
All segments referenced by a sequence must be placed in the
volatile playback memory of the baseband.
629 File format invalid; Unable to view/edit selected sequence file.
Single Waveform Sequence files cannot be viewed or edited at this
time. {name}
This error is reported when an single waveform sequence is selected
to and the user tried to view or edit the sequence from the front
panel. {name} is the single waveform sequence filename
630 Arb Sequence Error; Failed to write Arb sequence to sequencer.
This is reported when there is a failure in playing the selected
sequence.
630 Arb Sequence Error; Sequence contains too many levels of sequences
containing other sequences. Maximum nesting level is 32.
This is reported when an Arb sequence contains too many levels of
sequences within sequences. If a sequence contains another
sequence, that is one level of nesting. If that second sequence
contains yet another sequence, that is a second level of nesting.
The maximum depth of sequence nesting is 32 levels.
630 Arb Sequence Error; Sequence contains too many segments.
This is reported when an Arb sequence contains too many waveform
segments. The maximum number of segments that can be added to the
sequencer is not affected by the repetition counts of individual
segments.
630 Arb Sequence Error; Single wave segment length must be at least 60
samples. Value set to minimum.
This error is reported when the user attempts to specify a single
waveform sequence in which a segment is less than 60 samples. Due
to hardware limitations of the sequencer, segments must be at least
60 samples long.
630 Arb Sequence Error; Single wave segment offset must be even. Value set
to the next lower, even value.
This error is reported when the user attempts to specify a single
waveform sequence in which a segment begins at an odd sample
offset. Due to hardware limitations of the sequencer, segments must
Page 60