User manual

lpErrorDoc.txt
errors below 3.2 GHz.
The lowband coupler detector has failed. Report this error to
Agilent Technologies.
625 Internal pulse generator unlocked
*** No documentation ***
626 Internal modulation source unlocked
*** No documentation ***
627 Licensing error; A new challenge value must be generated using
'DIAGnostic:INFO:WLICense:CHALlenge?'
Most waveform license SCPI commands require an authentication code
which involves a challenge value. This challenge value must be
re-requested for each new command.
627 Licensing error; An unexpected time setting was detected. Time-based
licenses have been adjusted.
The internal clock is at a time which is earlier than the last
known setting. Time-based waveform licenses have been adjusted to
reflect the difference between the expiration date and the last
known time.
627 Licensing error; Because time-based licenses are present, setting the
date/time using SCPI commands is disabled.
The date/time SCPI commands are disabled when time-based licenses
are installed.
627 Licensing error; Can't license waveform since there are no 5-Pack
license slots free.
User attempted to add a waveform to a 5-Pack and there are no
5-Pack license slots free.
627 Licensing error; The specified authentication code is invalid
Most waveform license SCPI commands require an authentication code.
The specified code did not match the expected value.
627 Licensing error; The specified license ({name}) could not be installed:
{name}.
An error occurred during the installation of a license.
627 Licensing error; The specified license ({name}) could not be removed
because the license was not found.
An error occurred during the removal of a license.
627 Licensing error; The specified waveform license was not deleted because
it does not exist.
The user attempted to delete a non-existent waveform license.
627 Licensing error; User-provided challenge values cannot be zero.
The authenticated license existence query requires the user to
Page 58