User manual

lpErrorDoc.txt
During the system configuration scan of parallel bus boards, a
board with a bad configuration header was detected. Report this
error to Agilent Technologies.
617 Configuration Error; Bad board header in serial slot: {value}
During the system configuration scan of serial bus boards, a board
with a bad configuration header was detected. Report this error to
Agilent Technologies.
617 Configuration Error; Baseband generator control could not be created.
The baseband generator control creation process failed due to a
configuration file conflict or a hardware configuration error.
Report this error to Agilent Technologies.
617 Configuration Error; Configuration information file is invalid. Please
reinstall the file.
The CONFIG_INFO.TXT file was tampered with. This is a signal that
the file was modified, but the internal checks were bypassed.
617 Configuration Error; Found improperly configured devices.
An improperly configured device was found during the system
configuration scan. Report this error to the nearest Agilent
Technologies sales and service office.
617 Configuration Error; IQ modulations subsystem not created.
The hardware control creation process failed due to a configuration
file conflict or a hardware configuration error. Report this error
to Agilent Technologies.
617 Configuration Error; License file not found. Creating empty one.
The license configuration file is always present. If flash was
reformatted, this warning will be reported. It also means that no
license keys are currently present (since the file is new).
617 Configuration Error; No freq output - bad configuration.
The frequency control subsystem failed to be created due to a
hardware configuration problem. Although the user interface appears
normal, the instrument is not functional. Report this error Agilent
Technologies.
617 Configuration Error; No power output - bad configuration.
The power control subsystem failed to be created due to a hardware
configuration problem. Although the user interface appears normal,
the instrument is not functional. Report this error to Agilent
Technologies.
617 Configuration Error; Option {name} is invalid.
This error is reported when a software option is in the wrong place
in CONFIG_INFO.TXT. This signals that somebody tried to bypass
licensing.
617 Configuration Error; Sweep subsystem not created.
The sweep subsystem creation process failed due to a configuration
Page 56