One or more OCNS channels must be enabled for channel balancing to
occur.
600 RPP has tripped.
The reverse power protection circuit has been triggered. Repeated
tripping of this circuit can cause damage to the instrument.
601 Power search failed.
While executing power search, the level meter circuit failed to
return a meaningful value. This event indicates that the power is
in a range that the leveling loop cannot properly level. The power
will be set to the last properly leveled power.
605 DSP FW download failed
The instrument's firmware was unable to successfully initialize the
internal digital signal processor (DSP) on the CPU/Motherboard.
Report this error to Agilent Technologies.
605 DSP FW download failed; Opt 300 BERT capability.
The instrument's firmware was unable to successfully initialize the
internal digital signal processor (DSP) on the bit error rate
tester board for Option 300 capability. Report this error to
Agilent Technologies.
605 DSP FW download failed; Option UN7
The instrument's firmware was unable to successfully initialize the
internal digital signal processor (DSP) on the bit error rate
tester board (Option UN7). Report this error to Agilent
Technologies.
606 DSP times out; AnalogMod
*** No documentation ***
606 DSP times out; BERT flash memory access error.
The flash memory on the BERT board (Option UN7) failed to respond
within the appropriate amount of time. If cycling power does not
correct the problem, report this error to Agilent Technologies.
606 DSP times out; CPU/Motherboard DSP or BERT board failed to respond
within an appropriate amount of time. Contact service for help.
The digital signal processor (DSP) on the CPU/Motherboard or the
bit error rate tester board failed to respond within the
appropriate amount of time. Report this error to Agilent
Technologies.
606 DSP times out; Demod flash memory access error.
The flash memory on the DEMOD board failed to respond within the
appropriate amount of time. If cycling power does not correct the
problem, report this error to Agilent Technologies.
606 DSP times out; Demodulator communication failed.
The digital signal processor (DSP) on the demodulator board failed